AC_ERR = KH accelerator error
A.
B.
C.
6.
Please gently open the “Reaction Chamber” and check if the small white agitator is missing.
Please check if the pH probe are secured at proper depth.
Accelerator may be malfunctioned. Please contact your dealer.
pH<ERR = pH reading low error
A.
B.
C.
D.
E.
F.
7.
This error message will occur if the pH reading is below 4.5. Alarm will activate.
Please check if the pH probe is properly secured in the socket.
Please execute the pH probe calibration procedure.
Please gently open the “Reaction Chamber” and check if it is dry inside.
If the “Reaction Chamber” is dry, please check the water level sensor if it can move up and down
freely.
If all settings are correct, please contact your dealer.
pH>ERR = pH reading high error
A.
B.
C.
D.
8.
This error message will occur if the pH reading is higher then 8.8. Alarm will activate.
Please check if the pH probe is properly secured in the socket.
Please execute the pH probe calibration procedure.
If all settings are correct, please contact your dealer.
Page12
4.
Additional User Instructions
1.
2.
3.
Upon initial assembly and installation, please make sure to perform 5 (five) dKH readings to ensure optimal
performance and accuracy.
Upon using KHG, at any time if you wish to move KHG, you must use the "Purge Chamber" function to
completely drain out all liquids in the Reaction Chamber. Moreover, please detach and invert the Reaction
Chamber from the unit and pat dry the insides with paper towel. Failure to do so may cause the corrosive
salt water to spill and may damage the parts of KHG. In severe cases, the damage maybe permanent.
Only use KHG approved pH probe. Non-approved pH probe would result inaccuracy. If possible, please
always have a spare pH probe at hands ready to use. We recommend exchange the pH probe every three
months, and perform pH4 and pH7 calibration every time at the exchange. When removing a pH probe for
storage, you must thoroughly clean the probe, pat dry the surface. To store the pH probe, the glass bulb at
the tip of the probe must be immersed in 3M potassium chloride solution.