ELECTROMAGNETIC COMPATIBILITY INFORMATION
In IEC/EN 60601-1-2 4th edition. These limits are designed to provide reasonable protection against harmful
interference in a typical medical installation. This equipment generates, use and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause harmful interference
to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to other devices, which can be determined
by turning the equipment off and on, the user is encouraged to try to correct the interference by one
or more of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet on a circuit different from that to which the other device(s) are connected.
- Consult with the dealer.
Electromagnetic Emissions
The PIEZOART is intended for use in the electromagnetic environment specified below.
The customer or the user of the PIEZOART should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guide
RF emission – CISPR 11
EN 55011
Group 1
PIEZOART
uses RF energy for internal operation.
Therefore, its radiofrequency emissions are very low and
are not likely to cause any interference in nearby equipment.
RF emission – CISPR 11
EN 55011
Class A
Harmonic emission –
IEC 61000-3-2
EN 61000-3-2
Class A
Voltage fluctuations
Flicker emission –
IEC 61000-3-3
EN 61000-3-3
Compliance
PIEZOART
covers devices for usage in all establishments
other than domestic and that are not directly connected to
a low voltage power supply network, which supplies
domestic environment.
This is a class A product. In a domestic environment this
product may cause radio interference in which case the
user may be required to take adequate measures.
17
Summary of Contents for PIEZOART
Page 1: ...USER MANUAL ...
Page 23: ...MEMO 22 ...