Program Storage
1. Press the button to select user’s preferred program.
2. Finalize the setting by selecting the modes from the , , , buttons.
3. Press the button to save the settings.
■ SETTING
1) Volume Control
■ Error and Remedy
:
The following error codes are for those errors or issues that may lead the handpiece to stop.
Press the button to select the menu.
Press the button to increase the volume.
Press the button to decrease the volume.
button or button will lead to the main screen after saving the setting.
Press the button to select the menu.
Press the button to accept.
Press the button to deny.
Once user selected YES button, it will reset the setting.
If No button is selected, it will advance to the main menu.
Press the button to go back to main menu.
Cause
Error
Remedy
Foot Controller became damaged.
Pie
zo Handpiece cord is not connected properly.
Pie
zo Handpiece became
damaged.
Ultrasonic tip is not installed.
Ultrasonic tip is running by touching an object
.
Ultrasonic tip is broken.
Ultrasonic tip is not tightened properly.
Foot Controller cord is not connected properly.
Please contact the manufacturer.
Reconnect the Piezo Handpiece cord properly.
If the error doesn’t get solved out, contact the
manufacturer or dealer.
Please contact the manufacturer.
Install the ultrasonic tip.
Run the ultrasonic tip first then use for surgery.
Replace the ultrasonic tip.
Use the torque wrench to tighten the ultrasonic tip precisely.
Press the Foot Pedal once. If the error message does not disappear,
contact the manufacturer or dealer.
Reconnect the Foot Controller cord properly.
If the error message does not disappear,
contact the manufacturer or dealer.
E1
E2
E3
* In Programs 1~3, POWER and BOOST values are adjustable. Please refer to the setting values as described on "Tip Manual"
* Programs 4~5 have "FIXED" POWER and BOOST values and these are NOT adjustable.
Programs 4-5 are available to use for all Saeshin Piezo tips.
- Program 4 is for exclusive use of SP-020i
- Program 5 is for exclusive use of SP-028i
- When using a normal tip in Program 4~5, the power increases, but the tip may break.
2) Reset