25
SCIP Command Operation
6.5 System Commands
The following contains the SCPI system commands that are compatible with
Ethernet and USB control.
6.5.1 SYST
:
ERR?
Syntax
SYSTem:ERRor?
Description
Query the instrument’s error queue. A record of up to N errors is stored in
the instrument’s error queue. Errors are retrieved in first-in first-out (FIFO)
order. The first error returned is the first error that was stored. Each
additional error up to N is read by N subsequent queries (one for each
error). For this instrument N=20. The error queue has to be read until no
more errors are returned, otherwise the error status is not cleared.
Note: Some of the listed error codes are here for backwards compatibility
with legacy products and other are reserved for future applications. Not all
error codes are applicable to this Mini-Matrix.
Result
code, message
code: Numeric value with the error code (0 if no error).
Message: String with error message.
6.5.2 SYST:ERRALL?
Syntax
SYSTem:ERRALL?
Description
This command will return all of the errors in the error queue. Errors are
retrieved in first-in first-out (FIFO) order.
Result
code, message
code: Numeric value with the error code (0 if no error).
Message: String with error message.