64
internal rotation position. After reaching
this position, the cycle restarts: maxi‑
mum abduction and maximum external
rotation.
During operation the synchronized
mode is indicated in the top left‑hand
corner by the symbol
.
Non-synchronized:
Both motors run independently of each
other within the selected ranges of
motion.
To enable the non‑synchronized mode,
disable the synchronized mode.
During operation the non‑synchronized
mode is indicated in the top left‑hand
corner by the symbol
.
Default: synchronized mode enabled
m
Caution!
Patient hazard –
The synchronized operation is recom‑
mended as a general rule. However,
medical and/or therapeutic indications
may call for the non‑synchronized
mode. In the non‑synchronized mode,
particular caution and attention on
the part of the physician/therapist is
required to rule out any risk for the
patient.
n
Total therapy time
Under menu item ”total therapy time ”
you can view each patient‘s total ther‑
apy time (duration of all the patient‘s
therapy sessions).
To delete the stored therapy time:
Press and hold the
parameter key
for
5 seconds or select the ”New patient“
function.
n
Service menu
For service purposes only, refer to
Service Manual.
Reminder:
You save the selected parameter values by
pressing the
STOP
key.
5.4 Programming
ARTROMOT®-S3 Comfort
ARTROMOT®‑S3 devices of the Comfort
series offer two more programming levels
for additional functions.
The programming levels are selected in the
same way as on Standard models.
Programming levels 1, 2, 3 and 6 are identi‑
cal with programming levels 1 to 4 of the
Standard models.
All special functions are disabled in the
standard models.
The following treatment values, settings
and information can be entered/viewed on
the programming unit (1):
LEVEL 1:
– Abduction
– Adduction
– Internal rotation
– External rotation
LEVEL 2:
– Pause
– Therapy timer
– Speed
– New patient
LEVEL 3:
– Reverse on load ‑ motor A
– Reverse on load ‑ motor B
– Motor A ON/OFF
– Motor B ON/OFF
LEVEL 4:
– Warm‑up protocol
– Isolation program
– Therapy documentation
Abduction/adduction
– Therapy documentation
Internal/external rotation