Recommended separation distance
d = 1.2
√
P
d = 1.2
√
P 80 MHz to 800 MHz
d = 2.3
√
P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be less than the
compliance level in each frequency rangeb.
Interference may occur in the vicinity of equipment marked with the following symbol.
6
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measure field strength in the location in which the FL1401 is used exceeds the applicable RF compliance level above, the
FL1401 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the FL1401.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3.0 V/m.
NOTE 3: U
T
is the a.c. mains voltage prior to application of the test level.
Electromagnetic Immunity
FL1401 is intended for use in the electromagnetic environment specified below. The customer or the user of
the FL1401 should assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radianted RF
IEC 61000-4-3
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
< 5% U
T
(>95% dip in U
T
)
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
3 A/m
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
< 5% U
T
(>95% dip in U
T
) for 0, 5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
3 A/m
3 Vrms
3 V/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the FL1401
requires continued operation during power mains interruptions, it is recommended that the FL1401 be powered
from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF communications equipment should be used no closer to any part of the FL1401, including
cables, than the recommended separation distance calculated from the equation applicable to the frequency of the
transmitter.
19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU.qxd 4/22/11 5:43 PM Page 6
Summary of Contents for FLASH Lite 1401
Page 39: ...75 74 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 74 ...
Page 40: ...76 77 ...
Page 41: ...79 78 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 78 ...
Page 42: ...81 80 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 80 ...
Page 43: ...83 82 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 82 ...
Page 44: ...84 85 ...
Page 45: ...87 86 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 86 ...
Page 46: ...89 88 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 88 ...
Page 47: ...91 90 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 90 ...
Page 48: ...92 93 ...
Page 49: ...95 94 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 94 ...
Page 50: ...97 96 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 96 ...
Page 51: ...99 98 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 98 ...
Page 52: ...100 101 ...
Page 53: ...103 102 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 102 ...
Page 54: ...105 104 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 104 ...
Page 55: ...107 106 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 106 ...
Page 56: ...108 109 ...
Page 57: ...111 110 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 110 ...
Page 58: ...112 113 ...
Page 59: ...115 114 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 114 ...
Page 60: ...116 117 ...
Page 61: ...119 118 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 118 ...
Page 62: ...120 19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 120 ...
Page 63: ...19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 122 ...
Page 64: ...19_2497_051111_FLS_DFU_ML_19_2497_062306_FLS_DFU qxd 4 22 11 5 44 PM Page 124 ...