S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
4 2
1. Completely unzip the cover from the cushion.
2. Remove the foam insert from around the bellows, then
remove the bellows sheet .
3. Disconnect the faulty bellow strip from the tube
that connects it to the umbilical and
remove
the bellows. Apply the reverse procedure for bellows
replacement.
4. Spare bellows come in a set of 2 connected together
with tube which is glued in place, so when replacing
the bellows, you will always replace a set of 2.
BELLOWS 6
WAY T-VALVES:
PART NUMBER
10864
3.14 Cushion Bellows Replacement
1
B
B
A
3
A
2
4
1. Unzip the cover from the cushion and remove the foam
insert from around the bellows (see section 3.13).
2. Identify the tubes going to the umbilical and disconnect
the umbilical tubes from T-pieces. Unclip the inner cushion
clamp part from the outer cushion clamp part .
3. Disconnect the umbilical from the inner cushion clamp
part
. Apply the reverse procedure for umbilical
attachment.
UMBILICAL
ASSEMBLY
SHORT
CUSHION:
PART NUMBER
10832
3.15 Cushion Umbilical Replacement
1
3
2
A
B
A
B
A
A
Summary of Contents for QUATTRO Acute
Page 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Page 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Page 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Page 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Page 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Page 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Page 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Page 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...