When replacing the top label make sure that
all the buttons are in place before the new top
label applied.
LABEL TOP QUATTRO
T19R:
PART NUMBER 10563
MOU BUTTON
T19R:
PART NUMBER
10358
1. Disconnect the mattress umbilical from the
unit if it is still connected.
2. Deflate the mattress by releasing the CPR.
3. Unzip the top cover (there are two zips
at opposite corners of the mattress) and
remove from the mattress. Replace cover if
damaged.
SEE SPARES LIST FOR APPROPRIATE COVER
REQUIRED FOR MATTRESS.
4. Unfasten both the end poppers of the air
cell. To prevent the possibility of tearing the
polyurethane of the air cell it is important
to roll the popper off the socket rather than
applying a direct face on force.
3.10 Mattress Cover and Air Cell Replacement
2
3
4
1