8. Now select the “Data” tab and
then select “Set” from the drop
down field, and the information
will be sent to the unit, this will
take a few seconds. Then using
the up/down arrow
keys on the unit you can now
scroll through the information.
9. To clear any fault from the fault
log on the information screen,
select the “Clear faults” box, this
will take a few seconds.
10. Once the faults are cleared “OK”
will display above the “Clear
faults” box. Using the up/down
arrow keys on the unit
scroll through the information
menu to check all the faults have
been cleared.
S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
2 2
8
9
10