2. Faults and Warnings
1 3
Indication
Information
Check
Fault: Low pressure
System will alert if pressure
falls below minimum
allowable levels.
System continues to
operate.
Press MUTE to clear alarm.
• Check CPR is not open.
• Check O-rings on CCV assembly of unit for
damage.
• Press the mute button to clear the alarm.
• Note: alarm will reoccur if fault persists.
• Observe if warning is displayed again, and note
if one particular cell set (e.g.: 1, 5, 9, 13 …, or
2, 6, 10, 14…) is not inflating.
• Check cells for punctures or detached tube.
Replace cells as necessary.
Fault: Rotor system
System will not operate.
If this fault occurs, unit
will require opening and
assessment by a qualified
service engineer.
• Electrical connections to cycle control valve.
• Faces of rotor and stator.
• Failed opto slot sensor PCB.
• No mechanical damage to cycle control valve
assembly.
• Replace cycle control valve assembly.
Fault: Pump / Triac fail
System will not operate.
If this fault occurs, unit
will require opening and
assessment by a qualified
service engineer.
• Pump electrical connection not made.
Reconnect.
• Replace pump.
• Replace control board.
Warning: Battery
On board back up battery
not charging.
• Replace 3.6v mempac battery on control
board.
• Replace control board.
No power
on switch on
Either the IEC module is
damaged, or on board
500mA fuses on the control
PBC are blown.
• Replace 500mA fuses on PCB.
• Replace IEC module.
• Replace control board.
Warning: Uncalibrated
Do not operate system.
• Pressure sensor on control board needs
calibrating.
Fault: EMI fault
System will not operate if
this fault occurs.
• Unit has detected that the pressure sensor
amplifier is adversely affected by external RF
fields.
• Unit will require opening and assessment by a
qualified service engineer.
Summary of Contents for QUATTRO Acute
Page 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Page 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Page 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Page 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Page 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Page 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Page 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Page 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...