8. The unit display will light up and
the unit will begin to program.
9. Service data program will show
the progress (0% to 100%) as
indicated.
10. Service data display will then
show ‘Finished OK’ when
programming is complete as
shown. The unit will then beep
and start as normal.
S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
2 6
10
Time and Progress of programming will display
as PCB is programmed and field will display
“Writing…”.
9
8
11. Now turn the QUATTRO unit
off. You can now check that
the programming has been
successful by switching the unit
on with the information (data)
key depressed and scroll
through the menu with the up
arrow key until you reach
Software ID; this will now display
the program number that has
been uploaded.
11
HOLD
SCROLL
2.9 Testing Rotor and CCV Assembly
1. Apply the power lead to unit and switch the unit on, with the down
arrow key depressed at the same time. Once the unit has started,
release the down arrow key . This enables you to enter the test
menu. The unit will display ‘TEST MODE CAL’.
2. Press the down arrow key
to scroll through the test menu until
‘TEST MODE Rotor System’ is displayed.
3. Press the information (data) key and the unit will display ‘Rotor
System Finding Index’ while
the rotor finds its index
point, which is at 0 deg. This
will take 20 seconds.
1
2
HOLD
SCROLL
3
Summary of Contents for QUATTRO Acute
Page 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Page 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Page 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Page 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Page 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Page 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Page 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Page 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...