UltiMate 3000 Series:
SD, RS, and BM Pumps
Page
188
Operating
Instructions
Specification*
LPG-3400RS
DGP-3600RS
HPG-3200RS
HPG-3400RS
Operating Principle
Serial dual-piston pump
Compressibility
compensation
Fully automated, independent of the composition of the mobile phase
Flow rate rage
Recommended
Settable
mL/min
0.1 - 8
0.001 - 8
mL/min
0.05 - 8
0.001 - 8
Flow rate accuracy
±0.1%
Flow rate precision
< 0.05% RSD or < 0.01 min SD,
whichever is greater
Pressure range
2-103 MPa (290 - 15000 psi)
With a flow rate of > 5 mL/min, the pressure range decreases linearly down to 80 MPa (11.600 psi).
Pressure ripple
Typically <1% or <0.2 MPa, whichever is greater
Gradient formation
Low-pressure
proportioning
Dual low-pressure
proportioning
High-pressure proportioning
Proportioning
accuracy
±0.5% of fullscale**
±0.2% of fullscale
Proportioning
precision
<0.15% SD
Number of solvents
4
6 (2 x 3)
2
2 bzw. 2 aus 4
Gradient delay volume
Standard: 690 µL
(325 - 1,840 µL with optional mixing systems)
Standard: 200 µL
(35 - 1,550 µL with optional mixing systems)
Solvent degassing
Built-in 4-channel
degasser
External
(optional)
External
(optional)
Weight
13.6 kg
16.6 kg
16.3 kg
16.4 kg
Dimensions
16 x 42 x 51 cm (h x w x d)
PC connection
All functions controllable via USB (USB 1.1 or 2.0)
Integrated USB hub with three USB ports (USB 1.1 or 2.0)
I/O
Interfaces
2 digital inputs, 2 relay outputs
15-pin D-Sub port for connection of a Solvent Rack or degasser
Safety features
Leak sensor, active rear seal wash system, excess pressure monitoring
User input/display
LCD indicating system parameters
Standby button
3 LEDs (Power, Connected, and Status) for status monitoring
4 function keys for operation during initial installation and maintenance
GLP features
In Chromeleon: Support of automatic equipment qualification (AutoQ) and System Wellness monitoring.
All system parameters are logged in the Audit Trail.
Wetted parts
Stainless steel, titanium, zirconium oxide (ZrO
2
), ruby. sapphire, aluminum oxide
(Al
2
O
3
), PEEK™, PTFE, ECTFE,FEP, UHMW polyethylene, Kalrez
®
, Teflon
®
AF
(only LPG-3400)
Power requirements
100-120V, 60 Hz; 200-240V, 50 Hz; max. 150 VA
Emission sound
pressure level:
< 70 dB(A) in 1-m-distance
Environmental
conditions
Range of use: Indoor use; temperature: 10 °C to 35 °C (50 to 95°F); air humidity: 80% relative humidity,
non-condensing; overvoltage category: II; pollution degree: 2
* Typical operating conditions for measurable specifications (RS pumps): 1 or 2 mL/min @ 20 MPa or 40 MPa
** Proportioning accuracy: Typically ± 1.0% of fullscale for combinations other than AB
Summary of Contents for UltiMate 3000 Series
Page 2: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions ...
Page 4: ...UltiMate 3000 Series SD RS and BM Pumps Page II Operating Instructions ...
Page 5: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page III ...
Page 6: ...UltiMate 3000 Series SD RS and BM Pumps Page IV Operating Instructions ...
Page 7: ...UltiMate 3000 Series SD RS and BM Pumps Operating Instructions Page V ...
Page 8: ...UltiMate 3000 Series SD RS and BM Pumps Page VI Operating Instructions ...
Page 14: ...UltiMate 3000 Series SD RS and BM Pumps Page vi Operating Instructions ...
Page 24: ...UltiMate 3000 Series SD RS and BM Pumps Page 10 Operating Instructions ...
Page 58: ...UltiMate 3000 Series SD RS and BM Pumps Page 44 Operating Instructions ...
Page 74: ...UltiMate 3000 Series SD RS and BM Pumps Page 60 Operating Instructions ...
Page 116: ...UltiMate 3000 Series SD RS and BM Pumps Page 102 Operating Instructions ...
Page 136: ...UltiMate 3000 Series SD RS and BM Pumps Page 122 Operating Instructions ...
Page 200: ...UltiMate 3000 Series SD RS and BM Pumps Page 186 Operating Instructions ...
Page 204: ...UltiMate 3000 Series SD RS and BM Pumps Page 190 Operating Instructions ...
Page 228: ...UltiMate 3000 Series SD RS and BM Pumps Page 214 Operating Instructions ...
Page 236: ...UltiMate 3000 Series SD RS and BM Pumps Page 222 Operating Instructions ...