FI-521 User Manual
23
PRIM.U
kgF
LBF
Primary unit
Select the primary unit from kgf or
lbf. The second unit is lbf if kgf is
selected as the primary unit.
*
The calibration standard weight
must be in primary unit!
lbF
SECND.N
100-
125,000
10000
Second scale full scale
value
The division number under second
unit. The maximum is
1.25*(PRIM.N).
NEWTON.N
100-
125,000
10000
Newton unit
The division number under Newton
unit. The maximum is
1.25*(PRIM.N).
UNITS
0-6
6
Units key
0=kgf
1=lbf
2=N
3= kgf,lbf
4=kgf,N
5=lbf,N
6=kgf,lbf,N
MOTION
1-255
4
Motion window
1-255 = ±0.25d *(1-255)
OVER.LD
0-100
0
Overload display
limitation
0=FS+9d
1-100=101%FS -200%FS.
ABUSE
0-255
20
Set loadcell abuse level FS+(0%FS to 255%FS)
If the force has exceed this level, it's
considered to be abusive to the
loadcell
CONFIG
SubMenu1
SubMenu2
Option
Default
Parameter Description
Comment