FI-521 User Manual
19
4 Configuration
The Configuration Menu consists of seven different sub-menus. Within each sub-menu
are different menu options.
The config/cal switch must be set in the ON position in order to make changes to
specific parameters.
NOTE:
l
Each loadcell has its own CONFIG, CAL, IN.OUT parameters. Before
entering the Setup mode, select the loadcell to be configured with the
[CELL]
key.
l
The parameters of USER, MISC, and TEST are the same for all loadcells.
l
When "
Lo.VoL
" or "
Lo.BAT
" is displayed (the voltage to PCB is low),
CONFIG, USER, CAL, IN.OUT menus can’t be entered and edited.
4.1 Entering the Setup Mode
1.
Power on the indicator by pressing the
[ON/OFF]
key.
2.
Press the
[SELECT]
key for 3 seconds. The indicator shows "
Config
" to
indicate that you are in Configuration Menu mode.
4.2 Navigating the Main Menu
1.
Once in the “
Config
” menu, use the
[PRINT]
(up arrow) or
[CELL]
(down
arrow) key to move to the next available menu in the Configuration Menu.
2.
Use the
[ZERO]
key to access the desired menu.
3.
Use the
[PRINT]
(up arrow) or
[CELL]
(down arrow) key to choose a sub-
menu (parameter).
4.
Use the
[ZERO]
key to select the options within the sub-menu.
5.
Use the
[PRINT]
(up arrow) or
[CELL]
(down arrow) key to make a change.
Press the up or down arrow until the desired choice is on the display.
6.
Press the
[ZERO]
key to select the choice.
CONFIG
USER
CAL
MISC
TEST
IN.OUT
CELL