22
FI-521 User Manual
Table 4.2 CONFIG Menu Choices and Explanations
CONFIG
SubMenu1
SubMenu2
Option
Default
Parameter Description
Comment
CFG.ON
CFG.OFF
Seal switch position
The display will show whether the
seal switch is in the ON or OFF
position. This parameter can’t be
changed within the software.
RESET NO
NO
Reset
Config
menu
parameters to default
setting
YES
cell.ty
tension
Tension
Select whether the
connected loadcell is
used for tension or
compression force
Compres
DSP.POL
pos
POS
When CELL.TY=
Compres and calibrated
in negative direction
If DSP.POL=Pos, then reading will
be displayed in Positive.
If DSP.POL=Neg, then the reading
will be displayed in Negative
neg
PRIM.N
100 -
100,000
10000
Primary full scale
capacity
PRIM.D
0.0001
1
The division value under
primary unit
The division value under second
unit is automatically determined by
the indicator according to the
division value under primary unit.
0.0002
0.0005
0.001
0.002
0.005
0.01
0.02
0.05
0.1
0.2
0.5
1
2
5
10
20
50