FI-521 User Manual
4.12.1 Output Setting
1.
Enter the Setup Mode by pressing the
[SELECT]
key more than 3 seconds.
2.
Use the
[PRINT]
or
[CELL]
key to select IN.OUT>OUTPUT menu.
3.
Press
[ZERO]
to enter setting mode.
4.
Use the
[PRINT]
or
[CELL]
key to select MODE (setting output mode) or SPF1
or SPF2 (setpoint force) to set the corresponding data of the output.
5.
Press the
[ZERO]
key to confirm the choice.
l
When OUTx.yz is displayed, use the
[CELL]
key to change the output line
number(x).
m
Use the
[PRINT]
key to change level after power on (y) and after an
event trigger (z); x=1-2, y=0-1, z=0-1.
Normally, y/z=0 will make external relay close and y/z=1 will make the
relay open.
Y=0/1 means after the indicator power on, the output level is 0/1
z=0/1 means if current display force is over the setting data (SPF1/2),
the output level is 0/1.
m
Press the
[ZERO]
key to confirm the change.
l
When SPF1/2 is displayed, use the
[ZERO]
key to enter set-point1/2 data
mode.
l
When "UNIT.kgf/lbf/n" is displayed, use the
[CELL]
,
[PRINT]
, or
[UNIT]
key
to select unit of input force number.
6.
Next, use the
[CELL]
,
[PRINT]
, or
[UNIT]
key to input force number.
7.
Use the
[ZERO]
key to confirm and save the changes.
8.
In normal force measure mode, if SPF1 or SPF2 is set larger than 0, the
electronic level on output line will change according to the setting of OUTx.yz.
9.
Press the
[SELECT]
key to return to the last menu item.
10.
Press the
[ON/OFF]
key to exit this mode.