16
17
.
Clean the unit with a cloth dampened with water or mild detergent, then
wipe it dry with a dry cloth.
.
Wipe the electrodes using a cloth dampened with ethyl alcohol (75%
v/v).
.
Do not wipe the unit with benzene, gasoline, paint thinner, concentrated
alcohol, or other volatile detergents.
.
Do not place objects on top of the unit. It could damage the unit.
.
Do not disassemble the unit. Repairs should only be carried out by digiO2
authorized personnel.
.
There is no sensor in the ECG/EKG; therefore, this unit does not require
calibration.
.
Remove the batteries from this unit when you are not going to use it for a
long period of time (approximately three months or more).
.
Do not take measurements when there are drops of water on your skin,
such as from sweat or after bathing.
.
Do not take measurements where the unit will be exposed to strong
electromagnetic forces.
.
ln order to achieve accurate measurements, please replace new batteries
when low battery indicator is on.
.
Safety hazard may occur in case the final system does not provide
sufficient insulation to mains and earth, however, the equipment can be
safely connected to other IEC/EN 60601-1 approved equipment that has
all its connectors intended for connection to this unit separated from
earth by basic insulation.
.
Do not use this product in combination with non IEC/EN 60601-1
approved computers/adaptors.
.
Any ECG/EKG record erased from the unit cannot be recovered.
.
Do not wash this unit with water.
.
Do not store the unit in the following ambient conditions:
(1) Locations exposed to direct sunlight.
(2) Locations subject to high temperatures and high humidity.
(3) Wet or damp locations where water may get on the unit.
(4) Dusty locations.
(5) Near fires or open flames.
(6) Locations exposed to strong vibration.
(7) Locations exposed to strong electromagnetic fields.
.
Parts of electrodes and other connectors should not contact other
conductive parts including earth.
.
The medical system that will connect to this unit or another equipment
shall comply with IEC/EN 60601-1-1.
.
When ECG transfer data through USB cable to PC, the ECG is no power no
function, to avoid any elctronic shock, please do not use ECG to measure
when USB cable is connecting PC.
Maintenance and Storage
Summary of Contents for ECG-10X
Page 1: ...ECG EKG Recorder ...