To Clean the IR Temperature Sensor:
1. Gently place instrument upside down.
Figure 89
Place Instrument Upside Down
2. Visually inspect the IR temperature sensor to determine the type of
sensor installed.
3. For instruments with the flush sensor and brush bracket, the following
procedure can be utilized to remove particles from the brush. Carefully
place the head of the screw driver (10” length recommended) at the
bristles and use a sweeping side-to-side motion through the entire
length of the bristles no fewer than three times.
Figure 90
Cleaning the Brush Assembly
4. Select the appropriate cotton swab as described. Wet one end of the
cotton swab with 99% Isopropyl alcohol.
Summary of Contents for GAC 2500
Page 4: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 IV ...
Page 20: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 16 SECURITY ...
Page 24: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 20 INITIAL SETUP ...
Page 28: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 24 NAVIGATION ...
Page 66: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 62 SETUP ...
Page 88: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 84 GRAIN CALIBRATIONS ...
Page 90: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 86 DIAGNOSTICS ...
Page 102: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 98 MAINTENANCE ...
Page 104: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 100 TROUBLESHOOTING ...