OPERATOR’S MANUAL
GAC® 2500 Model (INTL)
11001-1655A-201603
TROUBLESHOOTING / 101
ERROR
CODE
ERROR
PROBABLE CAUSE
CORRECTIVE ACTION
1
Empty Cell Measurement
Out of Spec
Moisture or dirt buildup in cell.
Dump grain and verify cell is clean and free of
grain in and surrounding the cell.
2
Empty Cell Weight Out of
Spec
Empty cell measurement is out of tolerance
Dump grain and verify cell is clean and free of
grain in and surrounding the cell.
3
No Products Installed
No product calibrations were found on the machine.
Install product calibration files.
4
Fill Motor Jammed
The motor has stalled while loading grain from a
possible obstruction.
Clear blockage from cell.
5
Invalid Grain Calibration File
An invalid calibration file was selected.
Re-install the selected calibration file.
6
Moisture Too High
Measured product is above the moisture upper limit
of the calibration.
Verify sample filled the cell.
7
Moisture Too Low
Measured product is below the moisture lower limit
of the calibration.
Verify sample filled the cell.
8
Instrument Low Temp Limit
Exceeded
Instrument temperature is less than the allowed
limit of 0 degrees C.
Move unit to a warmer environment or allow to
warm up.
9
Grain High Temp Limit
Exceeded
Grain temperature has exceeded the unit
specification to analyze grain.
Allow grain to cool and then re-analyze.
10
Grain Low Temp Limit
Exceeded
Grain temperature has exceeded the unit
specification to analyze grain.
Allow grain to warm and then re-analyze.
11
Sample Weight Too High
Test weight is greater than the range specified for
the calibration selected.
Use a more representative sample and
re-analyze grain.
Verify correct grain calibration is selected.
12
Sample Weight Too Low
Test weight is lower than the range specified for the
calibration selected.
Use a more representative sample and
re-analyze grain.
Verify correct grain calibration is selected.
13
No Communication
An internal communication failure has occurred and
communication has been lost.
Power unit off and turn back on to reset.
14
Instrument High Temp Limit
Exceeded
Instrument temperature is higher than the allowed
limit of 45 degrees C.
Check unit ventilation or move to a cooler
environment.
15
Unit to Grain Differential
The difference between grain temperature and
machine temperature has exceeded the unit
specification to analyze grain.
Allow grain and machine temperature to equalize
and then re-analyze.
16
Internal Power Supply Out of
Spec
Unit internal voltage is out of specification.
Service is required.Contact DICKEY-john Europe
at +33 141 19 21 80 for service or DICKEY-john
US Technical Support at 1-217-438-3371.
17
Unable to Predict Moisture
A corrupt instrument calibration file or other
unexpected error.
Dump sample and re-analyze.
19
Instrument Needs Updated
Firmware in the instrument is out of date for this
application.
Update device with new firmware.
21
File I/O Error
An unexpected file i/o error has occurred.
An obstruction has occurred in the cell. Blockage
must be cleared before testing can resume.
23
Invalid GAC 2500 Model
Number
The instrument does not support software above
5.0.
Return instrument to DICKEY-john for service.
Contact DICKEY-john Europe at +33 141 19 21
80 for service or DICKEY-john US Technical
Support at 1-217-438-3371.
24
No Locale
A region name is not present in imported Region.ini
file.
Return instrument to DICKEY-john for service.
Contact DICKEY-john Europe at +33 141 19 21
80 for service or DICKEY-john US Technical
Support at 1-217-438-3371.
25
Could not Create Default
Locale
An error has occurred while attempting to create a
default locale.
Return instrument to DICKEY-john for service.
Contact DICKEY-john Europe at +33 141 19 21
80 for service or DICKEY-john US Technical
Support at 1-217-438-3371.
26
SD Card Missing
The SD card has been removed or is corrupt and
unable to load products for calibration.
1. Insert an SD card.
2. Verify existing SD card is good.
3. Contact DICKEY-john Europe at +33 141 19 21
80 for service or DICKEY-john US Technical
Support at 1-217-438-3371.
50
Weight Measurement Device
Error
An error has occurred with the load cell.
Power unit off and turn back on to reset.
Service unit if failure continues.
Summary of Contents for GAC 2500
Page 4: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 IV ...
Page 20: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 16 SECURITY ...
Page 24: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 20 INITIAL SETUP ...
Page 28: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 24 NAVIGATION ...
Page 66: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 62 SETUP ...
Page 88: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 84 GRAIN CALIBRATIONS ...
Page 90: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 86 DIAGNOSTICS ...
Page 102: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 98 MAINTENANCE ...
Page 104: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 100 TROUBLESHOOTING ...