3. OPERATION
Page 63
© 2008 DH Instruments, a Fluke Company
Enter the number of the unit position that you would
like to change. The display becomes:
Unit#5 1SI 2other
3altitude 4user
Select the desired pressure unit category (SI units include units
based
on SI units such as
mmHg), then select the desired unit from the unit menu.
<3altitude>
is not available in this
RPM4-E-DWT model.
The units of measure available are listed in Table 6.
Table 6.
UNIT function - available units of measure
<1SI>
<2OTHER>
<3ALTITUDE
>
<4USER>
<1Pa>
<2hPa>
<3kPa>
<4MPa>
<5mbar>
<6bar>
<7mmHg>
<8mmWa>
<1psi>
<2psf>
<3inHg>
<4inWa>
<5kcm2>
Not available
in this
RPM4-E-DWT
model
If
<4user>
is selected, the user unit must be defined. The display is:
1. Entry
field.
Define user unit:
1.000000 unit/Pa
Enter the number of user units per Pascal (Pa) in the entry field. Pressing
[ENT]
defines the
user unit and returns to the
<Set up unit #n>
screen.
The user defined unit label can be customized to any alphanumeric, four character label
using the remote command “UDU” (see Section 4.4).
See
Section
7.1.1 for the pressure unit conversion factors used by RPM4-E-DWT.
3.9.7
<7INTERN>
PURPOSE
To view, set, adjust, and maintain various aspects of RPM4-E-DWT’s internal operation.
OPERATION
To access the internal selections press
[SPECIAL]
,
<7intern>
. The display is:
1baro 2ReadRt 3RPT2x
4lo vnt 5log
The INTERNAL menu choices include:
<1baro>
View the real time output of the RPM4-E-DWT on-board barometer (see Section
3.9.7.1).
<2ReadRt>
Turn ON and OFF RPM4-E-DWT’s automated, rate of change dependent,
reading and integration and display rate feature (see Section 3.9.7.2).
<3RPT2x>
Feature not used in this RPM4-E-DWT model.
<4lo vnt>
Feature not used in this RPM4-E-DWT model.
<5log>
View the RPM4-E-DWT incident log (see Section 3.9.7.5).