E-DWT-10000-AF OPERATION AND MAINTENANCE MANUAL
© 2008 DH Instruments, a Fluke Company
Page 62
The cursor is on the current restriction level. Select a different level or
or
[ESC]
back to the MAIN RUN screen.
If no password is active
, selecting
<2edit password>
displays the user
password and allows it to be edited.
Password: pppppp
0 disables password
Once a password has been entered, the user level cannot be changed without
reentering the password.
Passwords can be up to six numbers in length and cannot start with a zero.
If
<0>
is entered, the password is made inactive and the user will not be required
to enter a password to access the user level menu. This condition, with a
security level of
<2low>
, is the factory default.
If there is an active password
, the
RPM4-E-DWT password entry screen
appears. The user must enter the user
defined password or the factory
secondary password to proceed further:
RPM4-E-DWT SNnnnn-
xx
Password: pppppp
The first field,
<nnnn>
, is the serial number of the RPM4-E-DWT, followed by a
second field,
<xx>
, that represents the number of times that a secondary
password has been used. The second field,
<xx>
, increments each time a
secondary password is used. The third field,
<pppppp>
is for user entry of the
normal password.
The
factory
secondary password is available in case the user’s password has
been misplaced or forgotten. It can be obtained by contacting a DHI Authorized
Service Center (see Table 25). The factory secondary password is different for all
RPM4-E-DWT’s and changes each time it is used.
3.9.6
<6PUNIT>
PURPOSE
To customize the selection of pressure units of measure available in the
[UNIT]
key menu.
PRINCIPLE
The
[UNIT]
function key makes available a choice of four default pressure units (psi, MPa,
kPa, bar) (see Section 3.7.3). RPM4-E-DWT also supports many commonly used units other
than those included in the default set up. These units can be made available for active
selection by customizing the UNIT function using
[SPECIAL]
,
<6Punit>
. This allows
RPM4-E-DWT to offer a very wide selection of units while simplifying day to day operation.
The typical user customizes the
[UNIT]
function key to support his/her most commonly used
units. Up to six units can be included in the
[UNIT]
screen.
OPERATION
To customize the
[UNIT]
function key, press
[SPECIAL]
,
<6Punit>
from the MAIN RUN
screen. The display is:
1.
Entry field to specify which unit position (1 to 6) of the
[UNIT]
function key menu is to be changed.
Set up unit #5