ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc.
Page 158
SYMPTOM
PROBABLE CAUSE
SOLUTION
PG7601 automated motorized rotation
is not operating.
Automated motorized rotation is OFF.
Press
[ROTATE]
,
<1mA>
PG7000
Operation and Maintenance Manual
PG7601 automated pressure control
(GEN) is not starting after setting a
target and loading mass AND
measurement mode is absolute by
vacuum.
In absolute by vacuum measurement
mode, before automated pressure
generation begins, the reference vacuum
must be below 20 Pa.
Wait for vacuum to reach less than 20 Pa,
check and correct vacuum supply if
inadequate.
Automated pressure control is not
increasing or decreasing pressure as
expected.
Pressure or vacuum supply on the
pressure controller is not adequate.
Provide correct pressure and vacuum
supplies to Control Cabinet. 6.13
PG7601 motorized rotation is unable
to start piston rotation.
Piston-cylinder is dirty.
Clean piston-cylinder. 6.6
PG7601 piston rotation rate slows
down too quickly; motorized rotation
engages too frequently.
Piston-cylinder dirty, PG7601 Platform not
level, PG7601 subjected to excessive
vibration, piston position indication needs
to be adjusted.
Clean piston-cylinder, level PG7601
Platform, remove sources of vibration,
calibrate piston position inidication. 6.6
PG7601 piston position readings seem
incorrect.
Piston position reading system needs to
be adjusted.
Execute piston position detection adjustment
procedure. 6.7
FPG8601 pressure indication is
unstable when the FPG8601 is
bypassed (vented).
Piston-cylinder needs to be cleaned.
Clean piston cylinder. 6.3
ADCS Tools displays
<FPG not
responding>
.
FPG8601 is in a condition in which it
cannot be used.
View in
ADCS Tools
FPG8601 Run Screen
.
Check for noisy output or other abnormal
conditions. 4.3.6.2
Pressure defined by ADCS-601 is
incorrect by relatively small amounts.
A pressure head is applied incorrectly or
inadvertently, the value of gravity used by
the system is incorrect, the PG7601
Platform is not level, information in the
piston-cylinder module and/or mass set file
is incorrect.
Check and correct if necessary. 4.2.6
When using automated pressure
control pressure generation stopped
and the controller is beeping.
Pressure reached controller’s UPPER
LIMIT.
Check
[GEN]
UPPER LIMIT and adjust if
necessary. PG7000 Operation and
Maintenance Manual
Automated pressure generation is
extremely slow or never floats the
pistons.
The test volume is too large.
Reduce the test volume
Summary of Contents for ADCS-601
Page 10: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page VIII N NO OT TE ES S ...
Page 12: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page X N NO OT TE ES S ...
Page 19: ...2 SYSTEM OVERVIEW Page 7 2005 DH Instruments Inc Figure 2 ADCS 601 system pneumatic schematic ...
Page 114: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 102 N NO OT TE ES S ...
Page 116: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 104 N NO OT TE ES S ...
Page 146: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 134 N NO OT TE ES S ...
Page 168: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 156 N NO OT TE ES S ...
Page 174: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 162 N NO OT TE ES S ...