6. GENERAL MAINTENANCE AND ADJUSTMENTS
Page 117
© 2005 DH Instruments, Inc.
6.4
PG7601, DISASSEMBLE PLATFORM AND REMOVE
MOUNTING POST PRT
It is not normally necessary to remove the mounting post PRT for recalibration. It can be recalibrated
in-situ by comparison with a reference PRT (see Section 7.3.4.).
To disassemble the platform and remove the mounting post platinum resistance thermometer, proceed as
follows:
Remove the piston-cylinder module from the mounting post (see Section 6.6.1) and replace with the
ORANGE storage plug.
Disconnect the PG7000 Platform from the PG7000 Terminal.
Invert the PG7000 Platform so that the bottom of the platform is up. Support the platform so that it
does not tip over. A simple solution is to place the platform upside down on a sturdy box which is
smaller than the outside platform dimensions and is also tall enough to allow the mounting post to
be suspended.
Remove the six socket head screws (3 mm) around the perimeter of the platform.
Lift the base cover plate gently and carefully disconnect the sensor leads at their board connections,
noting their locations. Remove the cover plate and electronic board attached to it.
Loosen -
but do not remove
- the four socket head screws (3 mm) located on the PRT/cable pass
through plate. The plate is located on the inside of the base.
Slide the PRT out of the mounting post.
To reassemble continue as follows:
After reinserting the PRT into the mounting post and tightening the four socket head screws,
reassemble following the steps above in reverse order. Thermal grease may be applied lightly to the
PRT, if available. If the sensor lead locations were not noted correctly, follow Table 19.
Power up PG7000 and verify proper operation of all on-board sensors.
Table 19.
PG7601
mounting post wire colors, description and location
WIRE COLORS
DESCRIPTION AND LOCATION
Black/Yellow/White RPM
(P4)
Black/Black/Black/Black Vacuum
Sensor (center raised micro board)
White/White/Yellow/Yellow
Mounting Post PRT (P2)
Purple/Blue/Red
Motor Solenoid (J2)
Gray/Orange/Green/Brown LVDT
(P1)
Summary of Contents for ADCS-601
Page 10: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page VIII N NO OT TE ES S ...
Page 12: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page X N NO OT TE ES S ...
Page 19: ...2 SYSTEM OVERVIEW Page 7 2005 DH Instruments Inc Figure 2 ADCS 601 system pneumatic schematic ...
Page 114: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 102 N NO OT TE ES S ...
Page 116: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 104 N NO OT TE ES S ...
Page 146: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 134 N NO OT TE ES S ...
Page 168: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 156 N NO OT TE ES S ...
Page 174: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 162 N NO OT TE ES S ...