ADCS-601™ OPERATION AND MAINTENANCE MANUAL
© 2005 DH Instruments, Inc.
Page 88
A Test Definition test has the test parameters and pressure sequence pre-
defined (see Section 4.4.1). To run a Test Definition test, a DUT Definition and
Test Definition must have been created (see Section 4.4.1.1, 4.4.1.2).
To run a Test Definition select
[Run]
,
[Run Test Definition]
. The run process
has three main steps:
Test
Initialization:
Steps through screens to:
- Select unit of measure. When running a Test Definition, the unit specified
by the Test Definition is always used to log data.
- Specify user.
- Select DUT(s) to be tested. Double click DUTs in the grid to add them to
the list of DUTs to be tested.
- Configure DUT(s). Specifies details specific to the DUT being tested.
- Select Test Definition to be run. By default, the Test Definition specified
in the DUT Definition is selected. Double click the desired Test Definition.
- After the Test Definition has been selected, most ADCS-601 users can
press
[Finish]
to skip remaining initialization steps and go directly to running
the test.
Run Test Sequence:
Runs leak test (if included) and pressure exercise (if
included) followed by the test points taking data at each point based on the
test data selections. The test parameters are defined by the Test Definition
(see Section 4.4.1.2).
End
Test:
Presents the Test Notes screen to record test specific notes and
the Test Complete
screen with various test conclusion options. Certain
aspects of the test conclusion can be customized using the
[Tools]
,
[Test
Options]
,
[End Test]
tab (see Section 4.4.1.4).
4.4.3
TEST RUN SCREENS
Special run screens are available when running an automated test.
The
Data Plot
and
Data Grid Run Screens
are available when running any automated test
for which there is a data file. These run screens visualize test results while the test is
running.
The
DUT/Reference Comparison Run Screen
is available when running any automated
test, regardless of whether there is a data file. This screen provides real time DUT and
ADCS-601 readings and information on the agreement between them.
See Section 4.4.3.1 for information on the
Data Plot Run Screen
, Section 4.4.3.2 for
information on the
Data Grid Run Screen
and Section 4.4.3.3 for information on the
DUT/Reference Comparison Run Screen
.
Summary of Contents for ADCS-601
Page 10: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page VIII N NO OT TE ES S ...
Page 12: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page X N NO OT TE ES S ...
Page 19: ...2 SYSTEM OVERVIEW Page 7 2005 DH Instruments Inc Figure 2 ADCS 601 system pneumatic schematic ...
Page 114: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 102 N NO OT TE ES S ...
Page 116: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 104 N NO OT TE ES S ...
Page 146: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 134 N NO OT TE ES S ...
Page 168: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 156 N NO OT TE ES S ...
Page 174: ...ADCS 601 OPERATION AND MAINTENANCE MANUAL 2005 DH Instruments Inc Page 162 N NO OT TE ES S ...