33
33
450182 • PU[REC] • Technical Reference Manual • Printing version 1.0.0 • November 2
nd
, 2020
MAIN SYSTEM
IEPE® via
MSI-BR-ACC
Input Range
± 10V
Sensor Excitation
4 mA ±10 %
Compliance voltage
>23 V
Accuracy
30 Hz to 30 kHz: 0.2 %
Power consumption
Max. 380 mW
Input coupling
AC 1.4 Hz
Bandwidth
70 kHz
limited by instrument
Signal-to-noise ratio; spurious-free SNR;
Effective number of Bits; Noise μV/V
PP
SNR
SFDR
ENOB
Noise
PP
Sample rate
[dB]
[dB]
[Bit]
[μV/V
PP
]
5 kS/s
101
124
17.1
0.88
10 kS/s
98
125
16.6
1.4
20 kS/s
83
123
14.1
1.9
50 kS/s
79
120
13.5
3.3
100 kS/s
76
115
13.0
4.5
200 kS/s
73
110
12.5
7
Sensor connection
BNC
TEDS
For adapter identification
MSI-BR-ACC
PU[REC]
Mode
IN+
IN-
V
IN
IEPE
10 V
4 mA
Range
Excitation
23 V
4 mA
220 nF
1 MΩ
1 MΩ
SHIELD
HOT
IEPE
®
sensor
MSI-BR-ACC