15
15
450182 • PU[REC] • Technical Reference Manual • Printing version 1.0.0 • November 2
nd
, 2020
MAIN SYSTEM
System specifications
PU[REC] - portable data recorder
>
16 analog input channels
(expandable to any signal input via Modular Smart Interfaces)
>
Quasi-static channel expansion via EPAD2
>
Available with 50 kS/s or 200 kS/s (optional) sampling rate
>
15.6” multi-touch display
>
Rugged and portable housing for easy transportation
PU[REC]
Configuration
Sampling rate / resolution
PUREC-50:
50 kS/s per channel
16-bit
PUREC-200:
50 kS/s to 200 kS/s
18-bit
100 S/s to 50 kS/s
24-bit
Digital input
2x counter shared with 8x digital inputs; 4x digital outputs
CAN bus
1x highspeed CAN 2.0 (ordering option PUREC-OPT-CAN )
Quasi-static channel expansion
EPAD2 interface connector
Expansion
SYNC-BUS (requires ordering option OXY-OPT-NET)
Main system
Display
15.6“ multi-touch TFT (full HD 1920 x 1080)
Additional Connectors
2x Display Port; 1x HDMI; Audio interface (3x 3.5mm connectors)
4x USB 3.0; 2x Gbit LAN;
Operating system
Microsoft Windows 10 64-bit; (optional Linux OS)
Data Storage
1 TB SSD in a removeable drive bay (870 GB useable for data storing)
up to 7 days of recording all channels at 50 kS/s or 300 days at 1 kS/s
MTBF
27800 hours
Noise emission
system idle
38 dBA
CPU max. heat; max. fan:
45 dBA
Dimensions (W x D x H)
463 x 129 x 318 mm (18.2 x 5.1 x 12.5 in.)
Weight
7.3 kg (16.1 lb.)
Power supply
Rated input voltage
100 to 240 V
AC
(max 90 to 264 V
AC
), active PFC
Input frequency
47 to 63 Hz
Maximal input current
2 A (230 V
AC
) / 4 A (115 V
AC
)
Inrush current
80 A (264 V
AC
)
Power consumption
max. 300 W; typical 65W (fully equipped with MSI, recording data)
Environmental specifications
Operating temperature
0 to +50 °C, down to -20 °C with prewarmed unit
Storage temperature
-20 to +70 °C
Humidity
10 to 80 % non cond., 5 to 95 % rel. humidity
Max. altitude
2000 m (6561 ft)
Sine vibration (EN 60068-2-6:2008)
Acceleration: 20 m/s²
Frequency range: 10 Hz - 150 Hz
Sweep: 1 oct/min
20 cycles
Shock (EN 60028-2-27:2009)
Acceleration: 15 g
Duration: 11 ms
Pulse form half sine
3 pumps/direction
6 directions
Random vibration (EN IEC 60721-3-2:2018)
Class 2M4
Spectral acceleration density: 1 m²/s³
Frequency range: 10 Hz-200 Hz
Duration: 30 min/direction
continued on next page ...