Error Codes
If INFO appears on the screen with a Code number, perform the corresponding Corrective
Action.
Code Description
Corrective Action
101
Received Signal Too Weak,
Measuring Time Too Long
Use the target plate or change the target surface.
102
Received Signal Too High
Target is too reflective. Use the target plate or
change the target surface.
201
Too Much Background Light
Reduce the background light on the target area.
202
Laser Beam Interrupted
Remove the obstacle and repeat the measurement.
301
Temperature Too High
Allow the device to cool down to a temperature
within the specified
Operating Temperature
Range
.
302
Temperature Too Low
Allow the device to warm up to a temperature within
the specified
Operating Temperature Range
.
401
Hardware Error
Switch the device on/off several times. If the error
still occurs, return the defective device to the
Service Center or distributor. Refer to the
Warranty
.
402
Unknown Error
Contact the Service Center or distributor. Refer to
the
Warranty
.
500
Data Error
Contact the Service Center or distributor. Refer to
the
Warranty
.