E7000L SitePROFILER
Handheld Cable/Antenna Analyzer
18
2-11
Test Panel Connector Overview
The connectors and indicators located on the test panel are described below.
External Power
The external power connector is used to power the unit and for battery charging. Input is 12.5 VDC to 19 VDC
at up to 3.42 A. The LED on the test panel indicates that the instrument is being charged up. This LED is colored
red when the battery is charging, and green when the battery is fully charged.
USB Interface - Type A
The E7000L has a Type-A USB port that accepts USB storage devices for transferring measurements, setup files,
screen images, GPS modes, and other data.
LAN port
The LAN port is used to connect the device to a PC with the SiteWorkbench software installed. This software
provides a simple and easy way to manage, archive, analyze, and print measurement reports, and customize
your cable list, antenna list, and signal standards list.
GPS port
With the GPS option, all measurements can be tagged and exported to geo-enabled maps for an accurate
record of site locations and measurements.
ECAL / Power Meter port
The ECAL/Power Meter port features a N(m) connector that serves to provide optional ECAL OSL calibration
and/or power meter measurements with the instrument’s built-in power meter and sensor.
RF In
50 Ω, N-Type female connector. This connector is used for ECAL and RF power measurements. Maximum input
is +30 dBm at ±50 VDC.
RF Out
50 Ω, N-Type female connector. This connector is used for Cable Analyzer reflection measurements. Maximum
input +25 dBm at ±50 VDC.
Optical Power Meter (OPM)
The OPM port features an FC connector, used with the optional Optical Power Meter (PN: DML-OPM) mode.
VFL (Visual Fault Locator) Symbol
The VFL port features an FC connector that provides an optional light source, aiding optical fiber continuity
checks and detection of spliced and/or bent fibers.