![DePuy Mitek Mini QUICKANCHOR Plus Manual Download Page 6](http://html1.mh-extra.com/html/depuy-mitek/mini-quickanchor-plus/mini-quickanchor-plus_manual_2485241006.webp)
6
2. Establish axial alignment of the QUICKANCHOR Plus
to the drill hole and push in to insert the Anchor.
WARNING: DO NOT TWIST OR APPLY BENDING
FORCE TO THE INSERTER AS EITHER MAY DAMAGE
THE ANCHOR, SUTURE OR INSERTER TIP.
WARNING: INCOMPLETE INSERTION OR POOR
BONE QUALITY MAY RESULT IN ANCHOR PULLOUT.
3. Retract the Slide Cover to the “Full Open” position
to release the Anchor, Suture and Needles
from the QUICKANCHOR Plus Handle and the
Suture Retaining Cap (FIGURE 1). Remove the
QUICKANCHOR Plus Tip from the drill hole and
gently pull back allowing the Suture and Needles to
feed out of the handle and paper carrier.
WARNING: DO NOT ATTEMPT TO REMOVE
QUICKANCHOR PLUS TIP FROM DRILL HOLE UNTIL
THE SLIDE COVER IS RETRACTED TO THE OPEN
POSITION.
4. Apply tension (approximately 8 lbs.) on the suture
lengths to set the Anchor in the bone.
WARNING: EXCESSIVE TENSION MAY OVERLOAD
THE ANCHOR OR SUTURE.
5. Complete the tissue reattachment.
STERILIZATION
The Depuy Mitek Mini QUICKANCHOR Plus is provided
sterile. Do not resterilize.
STORAGE
Store below 25°C (77°F), away from moisture and direct
heat. Do not use after expiration date.
Summary of Contents for Mini QUICKANCHOR Plus
Page 1: ...Mini QUICKANCHOR Plus P N 109169 Rev F Revised 09 12 2007 2012 DePuy Mitek...
Page 2: ...2 FIGURE 1...
Page 45: ...45 Bankart 1 2 3 DePuy Mitek 4 5 Depuy Mitek 6 7 8 Depuy Mitek ORTHOCORD...
Page 65: ...65 PDS Depuy Mitek Mini QUICKANCHOR Plus 0 2 0 1 2 3 DePuy Mitek 4 5 Depuy Mitek 6 7...
Page 66: ...66 8 Depuy Mitek ORTHOCORD Ti Al V Ni Depuy Mitek DePuy Mitek...
Page 87: ...87...