ENGLISH •  8

Ending the Day

1.  Remove ultrasonic/JET insert or air polishing nozzle 

from the handpiece, clean and sterilize.

2.  Place bottle with clean water in position “A.” Close 

Shield and set Selector Knob to position “A.”

3.  Operate ultrasonic scaler/JET at maximum water flow 

for at least 30 seconds to flush all previously used fluids 

from the system.

4.  Set selector knob to “Open•H

2

O” position and open 

Shield. Remove bottle with clean water from position 

“A” and place it in position “B”. Repeat steps 2 & 3  

in position “B”. Set Selector Knob to “Open•H

2

O” 

position, open Shield and remove bottle with clean 

water.

5.  Place an empty bottle in position “A” and close 

Shield.  Set Selector Knob to position “A” and operate 

ultrasonic scaler/JET for 30 seconds at maximum 

water flow to air purge the line. Set Selector Knob to 

“Open•H

2

O” position and open Shield.

6.  Remove empty bottle from position “A” and place 

it in position “B”. Repeat step 5 in position “B”. Set 

Selector Knob to “Open•H

2

O” position, open Shield 

and remove empty bottle from position “B”. At the end 

of the cycle, the line should be relatively dry.

7.  Clean and disinfect all surfaces of the DualSelect™ 

Dispensing System and attachments by applying an 

approved non-immersion type disinfectant solution 

carefully following the instructions provided by the 

disinfectant solution manufacturer. To clean system, 

generously spray disinfectant solution on a clean towel 

and wipe entire unit. Discard used towel. To disinfect 

system, generously spray disinfectant on a clean towel 

and wipe entire unit. Allow disinfectant solution air dry.

End of the Week

1.  Remove ultrasonic/JET insert or air polishing nozzle 

from the handpiece, clean and sterilize.

2.  Place bottle with clean water in position “A.” Close 

Shield and set Selector Knob to position “A.”

3.  Operate ultrasonic scaler/JET at maximum water flow 

for at least 30 seconds to flush all previously used fluids 

from the system. Set Selector Knob to “Open•H

2

O” 

position and open Shield.

4.  Remove bottle with water from position “A”  

and place it in position “B”. Repeat steps 2 & 3  

in position “B”. Remove bottle with clean water  

from position “B”.

5.  Place an empty bottle in “A” position, close Shield and 

set Selector Knob to position “A”. Operate ultrasonic 

scaler/JET for 30 seconds at maximum water flow to 

air purge the line. Set Selector Knob to “Open•H

2

O” 

position and open Shield.

6.  Remove empty bottle from position “A” and place it 

in position “B”. Repeat step 5 in position “B”. Remove 

empty bottle from position “B”.

7.  Prepare a 1:10 solution hypochlorite (NaOCl) solution 

Section  10 : System Care (con’d)

by filling 5.25% sodium hypochlorite to the designated 

line on the disinfectant bottle. Fill the remainder of the 

disinfectant bottle with clean water. Place disinfectant 

bottle in position “A”. Close Shield and set Selector 

Knob to position “A”. Operate ultrasonic scaler/JET 

at maximum water flow for at least 30 seconds until 

solution drains from the handpiece.

8.  Set Selector Knob to “Open•H

2

O” position and open 

Shield. Remove disinfectant bottle from position “A” 

and place it in position “B”. Switch Selector Knob to 

position “B” and operate for at least 10 seconds. Set 

Selector Knob to “Open•H

2

O” position, open Shield 

and remove disinfectant bottle.  

Leave unit undisturbed 

for at least 10 minutes but no more than 30 minutes to allow 

sodium hypochlorite solution to soak in lines. As a suggestion, 

PLACE A SIGN WHICH STATES THAT THE 

SYSTEM IS BEING DISINFECTED WITH A 

STRONG DISINFECTANT AND SHOULD NOT 

BE USED.

9.  Fill a bottle with clean water, place it in position “A” 

and close Shield. Set Selector Knob to position “A” 

and operate ultrasonic scaler/JET at maximum  

water flow for at least 30 seconds or until sodium  

hypochlorite odor disappears. ALL CHEMICALS 

MUST BE FLUSHED FROM THE SYSTEM 

BEFORE IT IS READY FOR PATIENT USE. Set 

Selector Knob to “Open•H

2

O” position and open 

Shield. Remove bottle from position “A” and place it 

in position “B.” Close Shield and set Selector Knob 

to position “B”. Operate ultrasonic scaler/JET at 

maximum water flow for at least 30 seconds until water 

drains or sodium hypochlorite odor disappears. No 

residual odor should remain in the system.

10. Set Selector Knob to “Open•H

2

O” position, open 

Shield and remove bottle with clean water from 

position “B”. Place empty bottle in position “A”, close 

Shield and set Selector Knob to position “A”. Operate 

ultrasonic scaler/JET system at maximum water flow 

for at least 30 seconds to air purge line dry. Set Selector 

Knob to “Open•H

2

O” position and remove empty 

bottle from position “A” and place in position “B”. 

Operate ultrasonic scaler/JET system at maximum 

water flow for at least 30 seconds to air purge line dry. 

At the end of the cycle, the line should be relatively 

dry.

NOTE: Dispose of all medicaments/disinfecting 

solutions from bottles on a weekly basis. Follow cleaning 

and sterilization procedures in the Cavitron Infection 

Control Procedure Booklet (Form No. 81158) provided. 

DISINFECTANT BOTTLE SHOULD BE USED FOR 

DISINFECTANT PURPOSES ONLY.

Section 10.2  Routine Maintenance

IMPORTANT: O-rings are a wear item and must be replaced 

routinely. Please contact your dealer or Cavitron Care

 for 

replacement Part No. 628051001.

Summary of Contents for Cavitron DualSelect

Page 1: ...e usar la unidad Bitte vor Inbetriebnahme des Ger ts sorgf ltig und vollst ndig durchlesen Leggere attentamente tutte le istruzioni prima di utilizzare l unit Dispositif de distribution DualSelect Mod...

Page 2: ...ntrol and Prevention and ADA American Dental Association guidelines by offering a closed line option and dispensing clean water Your DualSelect Dispensing System is UL Underwriters Laboratories listed...

Page 3: ...l Information 6 2 Unpacking the System 6 3 Water Line Requirements Recommendations 6 4 Water Supply Connection to the System 6 5 System Set Up 7 DualSelect System Component Description 5 7 1 System Co...

Page 4: ...ter is used close the dental unit water shut off valve at the end of the work day Fluid leakage may occur if Shield is opened with Selector Knob in A or B position Some medicaments may stain the syste...

Page 5: ...ine Connect the DualSelect water line blue to the external office water source line Open the manual shut off valve on the external water supply line and inspect all connections for leaks NOTE For clos...

Page 6: ...n Open H2 O position when removing bottles Cap Bottle B Bottle A Selector knob Selects between Bottles A and B and Open H2O which is external office water supply 7 2 Connections Back of System H2 O in...

Page 7: ...ap remove and place it upside down on a clean area to avoid contamination of the evacuation tube Fill bottle to the desired level and replace cap Tighten cap securely to prevent air leakage Insert bot...

Page 8: ...ose shield and set Selector Knob to appropriate position 8 Place a freshly sterilized insert nozzle into the hand piece 9 Unit is now ready for operation Between Patients 1 Remove ultrasonic JET inser...

Page 9: ...on B 7 Prepare a 1 10 solution hypochlorite NaOCl solution Section 10 System Care con d by filling 5 25 sodium hypochlorite to the designated line on the disinfectant bottle Fill the remainder of the...

Page 10: ...Product Specifications Dental water line pressure 25 60 psig 172 414 kPa Dental air line pressure 65 100 psig 448 690 kPa Output lavage pressure 30 psig 207 kPa Output air pressure 65 100 psig 448 690...

Page 11: ...uit ferm et en distribuant de l eau propre Votre syst me de distribution DualSelect est un produit inclus dans la liste UL Underwriters Laboratories Assistance technique Ce syst me ne contient aucun c...

Page 12: ...D ballage du syst me 6 3 Recommandations et sp cifications requises pour l alimentation en eau 6 4 Branchement du circuit d alimentation en eau au syst me 6 5 Mise en place du syst me 7 Description d...

Page 13: ...oduire si le couvercle de protection reste ouvert avec le bouton de s lection en position A ou B Certains m dicaments peuvent tacher le syst me et ou les flacons Pour viter la formation accrue d un bi...

Page 14: ...le syst me 5 3 Recommendations pour l alimentation en eau Le syst me de distribution DualSelect tout comme n importe quel dispositif d alimentation en eau pour cabinet dentaire doit tre install et ent...

Page 15: ...sont retir s Capuchon Flacon B Flacon A Bouton de s lection Choisit entre les flacons A et B et Open H2 O ce qui correspond l alimentation en eau du cabinet 7 2 Branchements panneau dorsal H2 O entr...

Page 16: ...e propre pour viter toute contamination du tube d vacuation Remplissez le flacon jusqu au niveau d sir et remettez le capuchon Serrez bien le capuchon pour viter toute fuite d air Ins rez le flacon en...

Page 17: ...uvercle de protection et retirez le flacon de la position B 7 Mettez le flacon contenant l eau propre le liquide m dicamenteux dans une position ou dans les deux fermer le capot et tourner le bouton d...

Page 18: ...e de protection 4 Retirez le flacon d eau de la position A et mettez le dans la position B R p tez les tapes 2 3 pour la position B R p tez les tapes 2 3 pour la position B Retirez le flacon d eau st...

Page 19: ...FLACONS DOIVENT ETRE UTILISES UNIQUEMENT POUR DES BESOINS DE DESINFECTION 10 2 Maintenance de routine Important les joints s usent et doivent tre remplac s r guli re ment Veuillez contacter votre repr...

Page 20: ...Prevenci n de Enfermedades de los Estados Unidos y de las ADA Asociaci n Dental Americana puesto que ofrece la opci n de suministro por v a cerrada y suministro de agua limpia Su sistema de suministro...

Page 21: ...3 Recomendaciones y requisitos para la red de distribuci n del agua 6 4 Conexi n de la red de distribuci n del agua al sistema 6 5 Preparaci n del sistema 7 Descripci n de los componentes del Sistema...

Page 22: ...gas de fluidos si se abre las cubierta protectora cuando la perilla de selecci n est en la posici n A o B Algunos medicamentos pueden manchar el sistema y o las botellas Para evitar que haya un aument...

Page 23: ...d completas de botellas desinfectante y t pon de almacenamiento 1 Paquete informativo 6 3 Recomendaciones y requisitos para la red de distribuci n del agua La presi n de la red de distribuci n del agu...

Page 24: ...la B Botella A Perilla de selecci n Permite seleccionar entre botella A y B y la posici n Open H2 O que es el suministro de agua del consultorio 7 2 Conexiones Dorso del sistema H2 O entrada Conecte l...

Page 25: ...9 1 Llenado de botellas Gire la perilla de selecci n a la posici n Open H2 O Levante la cubierta protectora lentamente y con cuidado hasta que quede totalmente abierta Al abrir la cubierta protectora...

Page 26: ...s posiciones Cierre la cubierta protectora y gire la perilla de selecci n a la posici n deseada 8 Coloque una boquilla recientemente esterilizada en el irrigador bucal 9 La unidad est lista para utili...

Page 27: ...a protectora gire la perilla de selecci n a la posici n A Haga funcionar el escarificador ultra s nico JET al flujo m ximo de agua durante 30 segun dos para purgar con aire la manguera Gire la perilla...

Page 28: ...81158 que le ha sido proporcio nado LAS BOTELLAS DE DESINFECTANTE DEBEN UTILIZARSE NICAMENTE PARA PROP SITOS DE DESINFECCI N Secci n 10 Cuidado del sistema cont 10 2 Mantenimiento de rutina Important...

Page 29: ...ndem die M glichkeit einer geschlossenen Leitung angeboten und reines Wasser abgegeben wird Das DualSelect Strahlabgabeger t entspricht den UL Richtlinien Underwriters Laboratories Technische Unterst...

Page 30: ...2 Auspacken des Ger ts 6 3 Anforderungen und Empfehlungen in bezug auf die Wasserleitung 6 4 Anschlu des Ger ts an die Wasserversorgung 6 5 Einrichtung des Ger ts 7 Beschreibung der Bauteile des Dual...

Page 31: ...B steht kann es zum Auslaufen von Fl ssigkeit kommen Einige Arzneimittel k nnen eine Verf rbung des Ger ts und oder der Flaschen verursachen Zur Vermeidung zunehmender Biofilmbildung wird die Verwendu...

Page 32: ...fektionsmittel und Verschlu zur Lagerung 1 Beutel mit Referenzmaterial 6 3 Anforderungen und Empfehlungen in bezug auf die Wasserleitung Der Druck in der Wasserleitung der Zahnarztpraxis mu zwischen 1...

Page 33: ...Flasche B Flasche A Wahlschalter W hlt zwischen Flaschen A und B und Open H2 O Open H2 O ist die externe Wasserversorgung der Zahnarztpraxis 7 2 Anschl sse R ckseite H2 O ein Blaue Wasserleitung des...

Page 34: ...ckung vorsichtig anheben und langsam in die vollst ndig ge ffnete Position bringen Das ffnen der Abdeckung l t den Flaschen berdruck entweichen Das Ger usch der entweichenden Luft ist als normal anzus...

Page 35: ...neten Position stellen 8 Einen frisch sterilisierten Einsatz oder eine frisch ster ilisierte D se in das Handst ck einsetzen 9 Das Ger t ist jetzt betriebsbereit Zwischen Patienten 1 Ultraschall JET E...

Page 36: ...Tuch spr hen und das gesamte Ger t abwis chen Desinfektionsmittell sung an der Luft trocknen lassen Kapitel 10 Pflege des Ger ts Forts serdurchflu wenigstens 30 Sekunden lang laufen las sen um die Le...

Page 37: ...ASCHEN D RFEN NUR DESINFEKTIONSMITTEL ENTHALTEN 10 2 Routinem ige Wartung Wichtig O Ring nutzen sich ab und mussen routinem ige ersetzt werden Zum Ersatz bon Teilent 628051001 setzen Sie sich bitte mi...

Page 38: ...rogando acqua pulita Questo sistema DualSelect approvato dalla UL Underwriters Laboratories Assistenza tecnica Nessun componente dell unit riparabile dall utente Delegare tutti gli interventi di assis...

Page 39: ...dazioni riguardanti il rifornimento idrico 6 4 Collegamento della rete di rifornimento idrico con il sistema 6 5 Installazione del sistema 7 Descrizione dei componenti del sistema DualSelect 41 7 1 Co...

Page 40: ...nit dentale alla fine della giornata di lavoro Si possono verificare perdite di liquido se il coperchio di protezione viene aperto quando la manopola di selezione in posizione A o B Alcuni farmaci pos...

Page 41: ...i dell aria e dell acqua e dispositivi di disinserimento rapido 2 Bottiglie e 2 tappi 1 Bottiglia di disinfettante e tappo 1 Opuscolo illustrativo 6 3 Requisiti e raccomandazioni riguardanti il riforn...

Page 42: ...tiglia B Bottiglia A Manopola di selezione Permette di selezionare tra bottiglia A e B e Open H2 O che la presa d acqua esterna dello studio 7 2 Prese di connessione Parte posteriore del sistema H2 O...

Page 43: ...hio lentamente e con cautela fino ad aprire completamente l unit L apertura del coperchio fa depressurizzare le bottiglie come si pu avvertire dal tipico rumore di sfiato Togliere le bottiglie come in...

Page 44: ...e la manopola di selezione nella posizione adatta 8 Inserire un ugello appena sterilizzato nel manipolo 9 L unit ora pronta per funzionare Tra un trattamento e l altro 1 Togliere dal manipolo l insert...

Page 45: ...lire l intera unit Lasciar evaporare la soluzione disinfettante 5 Porre una bottiglia vuota in posizione A e chiudere il coperchio Ruotare la manopola di selezione nella po sizione A e far funzionare...

Page 46: ...TTANTE DEVONO ES SERE USATE SOLO A SCOPO DI DISINFEZIONE 10 2 Manutenzione di routine Importante le guarnizioni O ring si logorano e vanno sostituite periodicamente Si prega di contattare il rappresen...

Page 47: ...ualSelect DualSelect DC Centers for Disease Control and Prevention ADA American Dental Association DualSelect UL Underwriters Laboratories Cavitron CareSM 1 800 989 8826 8 00 17 00 DENTSPLY DENTSPLY 1...

Page 48: ...47 1 48 1 1 1 2 2 48 3 48 4 48 5 49 5 1 5 2 6 49 6 1 6 2 6 3 6 4 6 5 7 DualSelect 50 7 1 7 2 8 51 8 1 8 2 9 51 52 9 1 9 2 10 52 53 10 1 10 2 11 54 11 1 11 2 12 54...

Page 49: ...48 1 2 3 DualSelect 1 200 2 1 1 1 2 DualSelect 8 2 4 10 10 CDC ADA 200 CDC ADA 2 5...

Page 50: ...SPLY 6 2 DualSelect 1 DualSelect 2 2 1 1 6 3 25 172 60 414 60 414 DualSelect 10 6 4 DualSelect DualSelect 6 5 Cavitron SPS Cavitron JET SPS Open H2 O DualSelect JET DualSelect DualSelect 5 5 1 10 5 2...

Page 51: ...50 7 DualSelect 7 1 A B Open H2 O A B Open H2O 7 2 H2 O DualSelect JET DualSelect DualSelect JET...

Page 52: ...51 8 8 1 1 2 2 2 3 DENTSPLY DENTSPLY 1 800 989 8825 717 767 8500 9 9 1 Open H2 O DualSelect 8 2 DualSelect 0 12 Cepacol Listerine 3 10 Viadent DualSelect...

Page 53: ...OPHY JET A Open H2 O B 180 Open H2 O A B B A JET DualSelect A B 30 Open H2 O Open H2 0 DualSelect A B 9 10 10 1 1 2 A 3 JET CDC 4 Open H2 O 5 B 2 3 B 6 7 8 9 1 JET 2 A 3 JET 30 CDC 4 Open H2O A B 2 3...

Page 54: ...H2O B 7 DualSelect 1 JET 2 A 3 JET 30 Open H 4 A B 2 3 B B 5 A A JET 30 Open H2O 6 A B 5 B B 7 NaOCl 1 10 5 25 10 A A JET 30 8 Open H2O A B B 10 Open H2O 10 30 9 A A JET 30 Open H2O A B B JET 30 10 Op...

Page 55: ...2 11 1 60 1 DualSelect 2 3 4 12 25 60 172 414 65 100 448 690 30 207 65 100 448 690 480 15 55 4 2 1 9 10 26 0 8 20 3 9 22 9 96 2 4 96 2 4 15 40 30 75 0 70 10 95 500 1060 5 6 11 2 Cavitron CareSM 1 800...

Page 56: ...mile Way York PA 17404 Phone 800 347 7412 Deutschland DENTSPLY DeTrey GmbH De Trey Strasse 1 D 78467 Konstanz Germany Phone 7531 583 0 United Kingdom DENTSPLY Ltd Hamm Moor Lane Addlestone Weybridge S...

Reviews: