
5 Commissioning
5.3 Putting the unit into operation
Dentsply Sirona
Instructions for Use Primeprint
46
67 45 173 D3691
D3691.201.01.01.02 04.2021
4.
Confirm the new network settings with
"Ok" or press "Cancel" in
order not to save the changed settings.
5.
If you have changed the setting to
"Automatic" you must now
disconnect the Ethernet connection to the PC and connect the
machine to the network socket in your local network.
6.
If you wish to change the settings back again, then change over to
"Manual IP settings".
IMPORTANT
If the machine is in
"Auto IP settings" mode and does not detect any
active DHCP server, the IP address is automatically reset to
192.168.230.1. With the help of a direct connection to a PC, you can
then add and manage the machine again via the
"Add Device
(Manual)" function.
Firmware download
Each CAD/CAM unit of the Dentsply Sirona requires a firmware version
that is compatible with the respective version of the inLab CAM
software.
You start the download of the appropriate firmware for Primeprint via
this button.
IMPORTANT
If a unit does not have the correct firmware version, this is in fact
detected by the software, yet it cannot be used for production. It is
indicated as "invalid firmware" both in the device management and in
the production phase.