7 Conditioning
Operating Instructions Midwest E Plus
7.3 Manual cleaning and disinfection
båÖäáëÜ
7.3
Manual cleaning and disinfection
Manual conditioning is possible in exceptional cases
IMPORTANT: Manual conditioning is possible in exceptional cases if the
national/regional regulations are followed. The national/regional
regulations are to be checked before.
Immediately after treatment
NOTICE! Condition immediately, or at the latest, one hour after treatment.
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
Using the T1 spray
NOTICE! Only use Midwest® Plus Aerosol Spray.
Use soft brush
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Procedure _ Instrument without integrated spray _ without Alpro
✔
Wear appropriate protective clothing.
✔
All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
1. Brush the instrument under running water (< 38 °C, < 100 °F, at least
drinking water quality) and good lighting (min. 500 lux) and color
rendering index (min. 80 Ra) until no more dirt can be seen, for at
least 10 seconds.
2. Flush the drive channels with spray.
3. Clean and disinfect spray channels with suitable agents and adapters
according to manufacturer's instructions.
4. Blow spray channels with suitable adapters out with 2.5 - 3 bar until
no more moisture displaces, but at least 10 seconds.
5. Conduct thermal disinfection or unwrapped steam sterilization.
6. Maintain mechanical parts manually [ → 15].
7. Maintain the push button chuck manually [ → 16].
8. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
9. Conduct sterilization [ → 17].
7.4
Manual maintenance
7.4.1
Maintenance of mechanical parts
Intervals
Daily at noon and in the evening
● At least every noon and evening
● Prior to each sterilization
after every thermal disinfection
● After every thermal disinfection without integrated maintenance
Required accessories
Required accessories
Spray adapter Midwest E Plus/ E Pro
Using the T1 spray
NOTICE! Only use Midwest® Plus Aerosol Spray.
B
A
A
Spray adapter
B
Midwest® Plus Aerosol Spray