Sirona Dental Systems GmbH
7 Conditioning
Operating Instructions Midwest E Plus
7.5 Sterilizing
båÖäáëÜ
7.5
Sterilizing
Interval _ US
Intervals:
● Prior to initial operation
● Prior to every other use
Process
Action steps_instruments
✔
The instrument is clean and disinfected.
✔
Instrument maintenance is complete.
✔
The instrument can be sterilized in packaging suitable for sterilization
and storage: paper/plastic composite packaging according to ISO
11607.
➢ Sterilize the instrument in the steam sterilizer with saturated water
vapor.
Unit pressure in bar and psi_US
without DAC PROFESSIONAL _ US
Gravity displacement steam sterilizers are permitted.
NOTICE! Do not exceed 140°C (284°F), even during the drying phase.
After sterilizing - heading
After sterilizing
Subsequent steps_instruments
1. Remove the instrument from the steam sterilizer immediately.
CAUTION! The instrument is hot. Risk of burns!
NOTICE! Do not attempt to accelerate the cooling process by
immersing the instrument in cold water. This can damage your
instrument.
2. Store all instruments so that they are protected from contamination.
3. Sterilize again once the storage period has elapsed.
Temperature:
135° C (275° F)
Overpressure:
2.13 bar (30.89 psi)
Article
Holding time at 135 °C
(275° F)
Drying time
Wrapped instruments 10 minutes
30 minutes
Unpackaged
instruments
3
.5
minutes
0 - 1 minute