![Dentsply Sirona Midwest E Plus Operating Instructions Manual Download Page 20](http://html1.mh-extra.com/html/dentsply-sirona/midwest-e-plus/midwest-e-plus_operating-instructions-manual_2483673020.webp)
66 04 206 D3686
16
D3686
.
201.01.01
.
02 08.2016
7 Conditioning
Sirona Dental Systems GmbH
7.4 Manual maintenance
Operating Instructions Midwest E Plus
Process
Maintenance _ Instrument
✔
The spray adapter is disinfected.
✔
The O-ring on the spray adapter is intact.
1. Fit the spray adapter onto the nozzle of the spray can.
2. Insert the instrument until it snaps into place and hold it.
3. Spray the instrument for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
4. Wipe any spray that comes out with a disinfection cloth.
5. Repeat the process until the spray is clear.
7.4.2
Care of the push button chuck
Sirona
New
Use Midwest® Plus Aerosol Spray on the push button chuck to remove
deposits and ensure proper functioning of the clamping system.
Interval
Interval
● At least once a week
Required accessories
T1 spray
Using the T1 spray
NOTICE! Only use Midwest® Plus Aerosol Spray.
Procedure
Vorgehensweise _ Winkelstück
✔
The instrument is clean and disinfected.
1. Press the contra-angle handpiece head with the chuck firmly against
the spray can nozzle.
2. Spray the chuck for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
3. Wipe any spray that comes out with a disinfection cloth.
A
B
A
Black nozzle insert
B
Midwest® Plus Aerosol Spray
Summary of Contents for Midwest E Plus
Page 2: ......
Page 4: ......
Page 5: ...k W l f MUKOMNS j b m b b Operating Instructions...
Page 27: ...k W k D MUKOMNS j b m c c Notice d utilisation...
Page 49: ...k W j MUKOMNS j b m b b Manual del operador...
Page 71: ......