6 Follow-up
Operating Instructions Midwest E Plus
6.1 After each treatment session
båÖäáëÜ
6
Follow-up
6.1
After each treatment session
Immediately after treatment
NOTICE! Condition immediately, or at the latest, one hour after treatment.
Using the T1 spray
NOTICE! Only use Midwest® Plus Aerosol Spray.
Instruments _ bur instrument
✔
The bur instrument must not be moving.
✔
Wear appropriate protective clothing.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Remove the bur instrument with tweezers.
3. Pre-disinfect directly at the treatment center [ → 14].
4. Detach the instrument from the motor.
5. Transport the instrument to the hygiene room in a suitable transport
container.
6. Conduct automatic conditioning [ → 14]. Manual conditioning [ → 15] is
possible in exceptional cases if the national/local regulations are
followed.
7. Apply spray to the instrument [ → 15].
8. Sterilize the instrument and accessories [ → 17].
6.2
At the end of the work day
Instrument
➢ Apply spray to the instrument [ → 15].
Remove the motor at night
NOTICE! Do not leave any instruments on the motor overnight, in order
to prevent oil from leaking into the electric motor. Never lubricate the
electric motor.