3. STEP-BY-STEP INSTRUCTIONS

3.1 digit power

Dispenser Installation: Traditional Style Delivery Unit

• Step 1: Test fit the Adapter post in the tool holder. If necessary, adjust the Adapter arm length by removing the screw and selecting the proper alignment for a customized fit.
• Step 2: Attach the Regulator to the dental hose via an ISO Threaded Connector or quick connect coupler depending on your configuration.
• Step 3: Place the Regulator in the tool holder to deactivate the air line.
• Step 4: Place the Adapter post in the tool holder to activate the air line. 

3.2 digit power

Dispenser Installation: Continental Style Delivery Unit

• Step 1: Remove Adapter arm by removing the screw. Loosen the threaded nut on the regulator, flip the regulator upside down and reattach 

it with the threaded nut. The Regulator knob should face down.

• Step 2: Attach the Regulator to the dental hose via an ISO threaded connector or quick connect coupler depending on your configuration.
• Step 3: Place the digit power

Dispenser on the tool rest to deactivate the air line.

• Step 4: To activate, pull hose forward and lay the digit power

Dispenser on the procedure tray.

3.3 Replacing digit power

Dispenser air supply hose

If a longer reach if desired, an extended Dispenser air supply hose is available separately. Shortening or replacement of a damaged air supply hose 
may be accomplished by following the Operations Manual directions below.

3.4 Quick Connect Regulator Conversion

The digit power

Dispenser Regulator is available with connections for standard ISO threaded screw connectors, MIDWEST

®

female quick connect/disconnect 

for Stylus

/XGT (does not fit MIDWEST

®

ATC), StarDental HiFlo

®

female quick connect/disconnect or KaVo MULTIflex

®

female quick connect/disconnect. 

Although not a routine maintenance procedure, the connections may be converted at the user’s discretion. See Operations Manual directions below.

3.5 Regulator use

• The Regulator speed control knob has four dispensing speed settings with 1 being the slowest and 4 being the fastest. 

The dispensing speed can also be set between numbers depending on your preference.

• Start out using the slowest speed setting and adjust upward as desired after becoming familiar with the technique.
• Check the pressure before each use and never bypass the Regulator.

3.6 Loading and using the digit power

Dispenser

To test functionality pre-operative, please follow directions below, extruding one cartridge, prior to use on patients.

• Prepare a digit power

Dispenser cartridge for use by snapping the circular disk off the cartridge. Insert unit dose cartridge into mixing tip. 

Ensure both cartridge spurs fully lock into the square holes on the side of the mixing tip. Use only the provided digit power

mixing tips.

• Remove the Cap from the Dispenser and insert the digit power

cartridge mixing tip assembly into the Cap. 

(Fig. 1)

• Lock the Cap onto the Dispenser with a ¼ clockwise turn until the alignment notches line up.
• To reduce Dispenser contamination, placing a digit power

protective barrier sleeve over Dispenser is recommended.

• Activate the Dispenser as outlined above. Depress the foot pedal and, away from the patient, bleed a small amount through the tip to assure 

proper flow and a good mix. 

(Fig. 2) See enclosed Directions for Use for complete material placement instructions.

• Start and stop the flow of material using the foot pedal.
• The dispensing rate can be adjusted by adjusting the Regulator speed control knob. 
• To disassemble, relieve pressure on the Dispenser plunger by stepping off of the foot pedal and place Regulator back into the tool holder. 

Unscrew Cap and remove. Remove used cartridge/mixing tip. Replace Cap prior to Dispenser disinfection.

• The blue plastic components – Cap, Grip and Adapter – can be autoclaved. Properly discard used cartridge/mixing tip and prepare 

Dispenser for reuse. (See Hygiene)

4. HYGIENE & MAINTENANCE

4.1 Cleaning the digit power

Dispenser

Remove protective barrier and discard. The digit power

Dispenser may be cleaned by scrubbing with a towel soaked with hot water and soap or detergent. Do not allow

solution to penetrate the casing. To clean plungers, remove Cap, depress foot pedal to advance plungers, and clean excess material with alcohol-moistened gauze.

4.2 Disinfection of the digit power

Dispenser 

Disinfect Dispenser with a hospital-level tuberculocidal disinfectant according to national and local regulations. 
• Sodium hypochlorite (0.525%), water based dual phenolics, alcohol based dual phenolics, and dual or synergized quaternary ammoniums are approved disinfectants. 
• Use impregnated wipes not sprays. Do not immerse the Dispenser in disinfectant.
The disinfectant manufacturer’s directions should be followed properly for optimum results. 

The blue plastic components – Cap, Grip and Adapter – can be steam autoclaved, at 132°C for a minimum of 5 minutes (unwrapped). Cap, Grip and Adapter may be
autoclaved at least 200 times. Use of other sterilization procedures/methods is at the discretion and sole responsibility of the practitioner.

4.3 Maintenance

No additional maintenance is required except for routine cleaning and disinfecting.

5. SERIAL NUMBER

The digit power

Dispenser and Regulator have different serial numbers located on their respective bodies. These serial numbers should be included in all correspondance.

Step 3

Step 1

Step 2

Step 4

Step 1

Step 2

Step 3

Step 4

Fig. 1

Fig. 2

Summary of Contents for digit Power

Page 1: ...d for input pressures of 60 to 80 psi Lower input pressures may affect output pressure and material dispensing rate Improper alignment of the notches on the cap and grip after loading a cartridge can...

Page 2: ...ridge into mixing tip Ensure both cartridge spurs fully lock into the square holes on the side of the mixing tip Use only the provided digit power mixing tips Remove the Cap from the Dispenser and ins...

Page 3: ...que tip Use your thumb and index finger to hold the push fitting down tightly while pulling the hose in the other direction Step 2 The new air hose can be cut to length with scissors if desired Step 3...

Page 4: ...into the bottom of the regulator using caution to avoid cross threading the parts Turn clockwise to install Step 6 Tighten the coupler insert using the 5 32 hex key Turn clockwise to tighten ORDERING...

Page 5: ...si Las presiones de entrada m s bajas pueden afectar la presi n de saluda y la velocidad de suministro del material Una alineaci n indebida de las muescas en la tapa y la empu adura luego de cargar el...

Page 6: ...r para usar desencajando el disco circular del cartucho Inserte el cartucho de dosis unitaria en la punta de mezcla Aseg rese de que ambos espolones del cartucho est n completamente trabados en los or...

Page 7: ...or mientras tira de la manguera Consejo t cnico Use el pulgar y el dedo ndice para presionar firmemente el racor mientras tira de la manguera hacia el sentido opuesto Paso 2 La manguera de aire nueva...

Page 8: ...r que desee en el fondo del regulador cuidadosamente para evitar enroscar las piezas en forma cruzada Gire hacia la derecha para instalar Paso 6 Ajuste el inserto del acoplador con una llave hexagonal...

Page 9: ...mauvais alignement des encoches sur le capuchon et la poign e apr s le chargement de la cartouche peut causer un mauvais fonctionnement de l appareil Il est important de toujours placer le distribute...

Page 10: ...ur Assurez vous que les deux perons de la cartouche se verrouillent dans les trous carr s sur le c t de l embout m langeur Utilisez uniquement les embouts de m lange digit power inclus Retirez le capu...

Page 11: ...pouce et votre index pour tenir le raccord instantan enfonc pendant que vous tirez la conduite dans l autre sens tape 2 Si vous le d sirez la nouvelle conduite peut tre coup e la longueur de votre ch...

Page 12: ...nsert de coupleur de votre choix dans le bas du r gulateur en faisant attention de ne pas croiser les filetages des pi ces Tournez la pi ce dans le sens horaire pour l installer tape 6 Serrez l insert...

Reviews: