- 15 /141 -
!
The
!
and
"
keys allow you to move to the right and left (continuous rotary systems only).
!
The
+
and
–
keys allow you to scroll through the file systems and files and to individually
change the torque or speed values in all continuous rotary file systems.
!
Motor settings cannot be changed for reciprocating systems.
!
When torque or speed values are changed from the default settings, the display will show
a
‘
in front of the torque or speed value.
6.4.3. Foot Pedal
The micromotor can be started in 2 alternative ways:
!
Either the foot pedal can be pressed and the micromotor operates as long as the foot pedal
remains pressed,
!
or the micromotor may also be operated by holding down the
OK
key
(7)
(See Fig. 4) for
1.5 seconds. The micromotor can be stopped again by pressing any key or the foot pedal.
To conserve battery power, when operated with the
OK
key, the micromotor will stop
automatically after 5 minutes of nonuse.
6.4.4. Acoustic Signals
When pressing the keys on the keypad, a series of audible tones and signals demonstrate that
the device is in use. All audible signals are active (unless changed manually).
All audible signals are activated upon delivery of the device:
!
Warning signal when torque exceeds approximately 75% of preset value in continuous
rotation.
!
Warning signal when torque exceeds a critical value in reciprocating motion.
!
Intermittent signal when in continuous counterclockwise rotation.
!
Warning signal before the device switches off when the battery is empty.
Deactivating and activating the acoustic signals
To deactivate the audible signals press the
!
and
"
keys simultaneously: a beep will sound
and the following message will be displayed:
To reactivate the audible signals, press the
!
and
"
keys simultaneously: a beep will sound
and the following message will be displayed:
Summary of Contents for WaveOne endo motor
Page 1: ...User Manual ...