- 18 /141 -
NOTE
The WaveOne
!
(MAILLEFER 6:1) contraangle reduces the speed so that the instrument will
rotate at the speed shown in the display.
During the calibration process the display will read:
The process automatically stops as soon as calibration has been completed, showing all black
bars on the display.
NOTE
Should you at any time wish to stop the calibration process, press any key or the foot pedal.
The display will read:
Possible Error Warnings
The display will then read for example:
Error 1:
The micromotor is not properly connected to the device or it is damaged. Check
the connection first.
Error 2:
The contraangle has a high resistance during rotation. Check the contraangle.
Lubricate if necessary.
6.5.3. A.S.R.: Automatic Stop Reverse
NOTE
ASR function only works in continuous rotation. In reciprocating mode, ASR is disabled (LED is
off).
The ASR function can be changed by pressing the
A.S.R. (9)
button:
(see Fig. 4) and the LED color will change according to the selected function: