8.
Troubleshooting
If your WaveOne
!
endo motor does not seem to work properly, review the checklist below.
If the problem persists, contact your distributor.
Problem
Possible Cause
Solution
THE DEVICE
DOES NOT
OPERATE
PROPERLY
The battery is not charged.
The battery charger is not plugged into
the power socket properly.
Network voltage does not correspond
to the voltage indicated on the battery
charger label when battery is being
recharged.
!
Charge the battery.
!
Check that the battery charger is plugged in
properly.
!
Check that the original battery charger is used.
!
Reload the factory default parameters.
THE DISPLAY
DOES NOT
OPERATE
PROPERLY
The display may be instable or tends to
dim as the battery charge is low.
!
Charge the battery.
THE
MICROMOTOR
DOES NOT START
Either the micromotor is improperly
connected to the housing or the contra
angle is malfunctioning.
!
Check that the micromotor connector is properly
inserted in the micromotor housing.
!
Check that the contraangle operates correctly.
!
Remove the contraangle and set maximum
speed, then start the micromotor again.
!
Calibrate without the contraangle, then
reconnect the contraangle and run calibration
again.
THE FOOT PEDAL
DOES NOT START
THE
MICROMOTOR
Either the foot pedal is damaged or it is
not plugged in properly.
!
Start the micromotor by pressing the
OK
key (7)
(See Fig. 4) for 1.5 seconds. If the micromotor
starts, first check if the foot pedal is connected
properly to the device. If yes, contact your
Service Center for foot pedal replacement.
THE BATTERY
DOES NOT
OPERATE
CORRECTLY
Battery discharges too rapidly even
though all precautions have been
observed.
The device only operates when the
battery charger is connected to the
mains supply, but does not operate with
battery power.
!
Battery might be damaged. Send the device to
your Service Center.
CALIBRATION
ERROR 1
An improperly connected micromotor
might disturb the calibration process.
!
Check that the micromotor is correctly connected.
CALIBRATION
ERROR 2
The calibration process could be
disrupted by an excessively resistant
contraangle.
!
Check the contraangle for any malfunction.
!
Lubricate the contraangle carefully.
Continuous rotary
file blocks in canal
Wrong file setting.
Too much pressure on the instrument.
!
Switch to ASR mode “Reverse” (LED red), start
the motor and pull out the file carefully.
Reciprocating file
blocks in canal
Too much pressure on the instrument.
File not frequently cleaned.
!
Try to remove the file with a pair of pliers by
pulling out and rotating the file gently clockwise.