- 12 /141 -
WARNING
If the BATTERY LED starts flashing red during use, connect the device immediately to the
battery charger, as the device might switch off.
It is possible to use the WaveOne
!
endo motor while it is charging.
If the battery is low and the device powers off, the device must be connected to the AC power
supply to continue the procedure.
Flashing orange:
Battery charger is correctly connected to socket
A
(see Fig. 2) on the back
of the device and battery is charging.
If the device powers off and the battery charger is connected, the unit will switch back on and
the following display reading will appear:
When charging is complete, the battery LED switches to green.
WARNINGS
!
Should any liquid exit the device that may be attributed to battery leaks, interrupt work
immediately and send the device to an authorized service center for battery replacement.
!
It is dangerous to open the device for battery replacement. Opening the device will
invalidate the warranty.
NOTES
!
In order to ensure good battery life, it is recommended to always work on battery power
and recharging the battery only when it is fully discharged.
!
If the battery is charged correctly, the operating time between two charges is a minimum of
approx. 2 hours.
!
Charging the battery completely may require up to 3 hours.
!
For optimal battery performance, the battery should be replaced every 2 years by an
authorized service center.
!
Upon charging, the device can be used normally and without appreciable increase in
charging time as the WaveOne
!
endo motor battery charger is sufficiently powerful to
directly power the micromotor while charging the battery.