MONTHLY MAINTENANCE PROCEDURES (CONT.)
Replenisher System (9000 Only)
1.
Empty, rinse and fill the Replenisher Reservoirs half full with warm water.
2.
With the Solution Trays removed, turn the Main Power Switch ON. The pumps will
run
approximately 6-9 seconds. Turn the Main Power Switch OFF, wait for
about 15 sec-
onds and then turn back ON for another 6-9 second pump cycle.
Continue this proce-
dure until the reservoirs are empty. Let the pumps pull
air the last time to clear them
and the lines of water.
NOTE:
Direct the stream of water from the pumps into the sump of the processor.
3.
Turn the Main Power Switch OFF and reload the Replenisher Reservoirs with fresh
chemistry.
4.
Before starting the machine at the beginning of the next day of processing, replace
both Transport Modules in their original solution trays.
Do not interchange posi-
tions.
5.
Fill the Solution Trays with fresh solutions until the solution level reaches the top of
the
overflow spouts. Turn the Main Power Switch ON.
CONTROL MODULE
There is no required user maintenance for the Control Module. If the processor fails to operate,
or for removal of electrical components, the following receptacles are located on the left side
of the Control Module. Verify that each component is properly connected.
Motor - 6 Pin Plug
No color code
Heater Pad - 2 Pin Plug
Tan or Beige
Solution Temperature Sensor - 4 Pin Plug
Orange
Dryer Heater & Fan - 3 Pin Plug
Pink
Water Solenoid - 3 Pin Plug
Light Blue
Replenisher (9000 Only) - 5 Pin Plug
Yellow (Capped off on 810 Ba-
sic)
Film Sensor Bar (9000 Only) - 9 Pin Plug
No color code (Capped off on 810
Basic)