ON/OFF POWER SWITCH AND ELECTRICAL CONNECTORS
The On/Off Power Switch is located at the rear of the Control Module and is accessible through
the rear Access Cover. All electrical connectors and the Power Cord are located toward the rear
of the unit for easy access.
CASE AND ACCESS COVERS
The molded Case is water-tight and has a drain connection for the waste effluent. The two
Access Covers can be removed for solution loading, maintenance and cleaning.
Figure 3: 9000 and 810 Basic Control Panels
17
CONTROL MODULE WITH CONTROL PANEL
The processor's Control Module is mounted within the machine beneath the Rear Access Cover.
It incorporates all electronic components necessary for the complete functioning of the system.
The module's housing and cover are molded of corrosion- proof plastic which protects the
electronics from the chemical environment. Located in the front of both the 810 Basic and 9000
Control Modules is a Control Panel which incorporates the following:
1. A sequential three-speed processing Time Selector.
2. Three lights directly above the Time Selector for indicating the throughput time selected.
3. A Processing light which indicates when the machine is in the process mode.
4. A Low Temp light which indicates that the temperature of the developer is below optimal
processing temperature (83 degrees F / 28 degrees C).
5. A Film Present light (9000 model only) which indicates that a film is detected by the film sensor
bar located above the film entry slot at the front of the machine. There is a 13-15 second delay
from the time film clears the sensor bar to when the light goes out. This serves as a "Wait" light
for helping to prevent film overlapping or light fogging.
6. A Run button (810 Basic model only) which puts the machine into and out of the processing
mode.
810 Basic
*+,&
26'
!
"#$
%
&'()
02)4&33+'5
-)./*&,0
9000
!
"#$
%
&'()
*+,&
-)./*&,0
1+-,/02&3&'*
02)4&33+'5