57
ELITE
pro
XC End View of CT and Voltage Lead Connections
2) Apply the CTs to the phases being monitored observing proper CT orientation.
3) If there is an arrow on the CT label, the label should point towards the load. Otherwise, use the
instructions printed on the CT, for example: “This Side Toward Load.”
4) Connect the voltage leads to the different phases. Connect the
Neutral wire first, then the remaining
voltage sense (phase) wires.
ELOG
Q
UICK
S
ETUP
W
IRING
D
IAGRAMS
When a Quick Setup is selected, the
View Typical Setup
button appears in the Quick Setup area of the
screen. Clicking this button displays a wiring diagram that corresponds to the selected setup, showing
how to connect the CTs and the voltage leads of the
ELITE
pro
XC
. The diagram can be printed and
carried into the field for easy reference during installation.
The following pages show the diagrams available for each Quick Setup.
CT Connections
Voltage Lead
Connections
www.
.com
1.800.561.8187
Summary of Contents for ELITEpro XC
Page 58: ...58 Typical Single Phase 2 Wire Setup Diagram www com information itm com 1 800 561 8187 ...
Page 59: ...59 Typical Single Phase 3 Wire Setup Diagram www com information itm com 1 800 561 8187 ...
Page 60: ...60 Typical 3 Phase 3 Wire Delta Setup Diagram www com information itm com 1 800 561 8187 ...
Page 61: ...61 Typical 3 Phase 4 Wire Wye Setup Diagram www com information itm com 1 800 561 8187 ...