The instrument is not suitable for aggressive acids
and bases. Vapours could destroy pump.
Safety instructions during and after use
`
`
The instrument must only be used within the
confines of applicable material resistance and for its
designated purpose. Any other type of use will affect
the warranty/guarantee.
`
`
Do not store the instrument with filled sample bottle.
`
`
The instrument is not suitable for cleaning in a
dishwasher.
Layout and function
`
`
The instrument is made of conductive material and
is suitable for filling non-aggressive, combustible
liquids (group IIC, see annex) from a maximum
suction depth of 5 m.
`
`
The instrument is suitable for sampling liquids from
0 - 60 °C.
`
`
The instrument is only suitable to take samples and
not for continuous operation.
`
`
Only use the instrument with the recommended
sample bottles specified in the accessories.
`
`
Protect the instrument against effects of weather
exposure during storage.
`
`
The instrument is maintenance-free.
Parts in contact with medium
`
`
PA, electrically conductive (suction hose)
`
`
Glass (sample bottle)
`
`
EPDM (head gasket)
`
`
Brass, nickel-plated brass (Hose weight)
Parts not in contact with medium
`
`
Aluminium (pump head, pump cylinder, frame)
`
`
NBR, EPDM (gaskets)
`
`
PA, electrically conductive (handle)
Scope of delivery
`
`
Check delivery for completeness and transport
damage.
`
`
Point out any damage in the packaging to the
freight forwarder/supplier immediately.
Assembly
1. Screw the sample bottle
onto the instrument
finger-tight.
2. Guide the suction tube
through the open clam-
ping screw
about 1-3 cm into the sample bottle.
3. Screw on the clamping screw finger-tight and fix the
suction tube in place.
4. Place the free end of the suction tube in the liquid
you are sampling. Ensure that the suction hose is
not kinked.
WARNING!
Possible sparking due to electrostatic discharge.
Danger of explosion if combustible liquids are
used.
`
`
Connect the earthing terminal
before use.
`
`
Wear the necessary protective clothing.
NOTE!
Before using, the instrument has to be cleaned to
the extent of your (validated) cleaning process.
`
`
Measures of manual cleaning have to be carried
out in the cleaning/desinfection area and under
consideration of occupational health and safety.
`
`
Pay attention to the material resistance.
CAUTION!
Liquid entering the instrument due to overfilling
or tipping. Bodily injury due to escaping liquid and
damage to the instrument.
`
`
Shut down and secure the instrument immedi-
ately.
`
`
Contact the dealer or manufacturer.
1 x
Aluminium ProfiSampler with frame
1 x
Sample bottle with closure, 500 ml
(glass, GL45 thread)
5 x
Suction tube (PA, electrically conductive,
black)
1 x
Tube weight (brass, nickel-plated brass)
1 x
Clamping screw
1 x
Earthing cable, length 2m
1 x
O-ring (EPDM)
1 x
Hose cutter
1 x
Directions for use
20 x Seal-it lead bottle seals
WARNING!
Danger of cutting when using the hose cutter.
`
`
Keep hands away from the cutting area.