Carbon dioxide

 

  

Detection method

 

Non

-

dispersive infrared 

(automatic calibration)

 

Operating range

 

350

-

10000 ppm

 

Precision

 

± 30ppm ± 3% 

 

Temperature

 

 

 

Detection method

 

Semiconductors

 

Operating range

 

0

-

50 °C (32

-

122°F)

 

Precision

 

± 1.0 °C (± 1.8 °F)

 

Humidity

 

 

 

Detection method

 

Semiconductors

 

Operating range

 

-

 95% rRH 

 

Precision

 

±5% at 20

-

80%rH 

 

@ 15

-

30°C (59

-

86°F) 

 

Fine/Suspended particles

 

 

 

Detection method

 

Laser diffraction

 

Operating range

 

0

-

1000 

μ

g/m3

 

Precision

 

>100 

μ

g/m3 , ± 10 % 

 

<100 

μ

g/m3 , ± 10 

μ

g/m3

 

V. Product technical parameters

 

Item

 

Content

 

Product dimensions

 

142x68x42 mm

 

Terminal block

 

Maximum of 1.5 mm

2

 

Rated voltage

 

9

-

24VDC +/

-

10% Or use 12V1A power supply 

 

Power consumption

 

5W

 

Operating temperature

 

0°C–

 

50°C 

 

Operating humidity

 

0%RH –

95%

RH (Non

-

condensing)

 

Storage

 

-

20°C–

 

+70°C 

 

Protection level

 

IP20

 

Certification

 

NCC,FCC/IC, CE, JRF

 

Accessories

 

Wall mount x 1, power supply x 1 (optional), external temperature sensor cable x 1 (optional), 
wall plugs and wall screws, hex head screws, terminal male connectors, serial number sticker x 1 

 

VI. Sensor parameters

 

Total organic compounds

 

 

Detection method

 

Metal oxides

 

Operating range

 

0

-

60000ppb

 

Precision

 

± 

15% (Data tested in 

laboratory using alcohol)

 

Formaldehyde

 

 

Detection method

 

Electrochemistry

 

Operating range

 

0

-

5000 ppb

 

Precision

 

±15% @20

-

50°C 

 

Carbon monoxide

 

 

Detection method

 

Electrochemistry

 

Operating range

 

0

-

500 ppm

 

Precision

 

±20 ppm

 

Ozone

 

 

Detection method

 

Electrochemistry

 

Operating range

 

0.4

-

10 ppm 

 

Brightness

 

 

Operating range

 

0

-

10000 lux

 

Summary of Contents for UNO-C07X011

Page 1: ...1 UNOnext Indoor Air Quality Monitor UNOC07X011 Packaged user manual ver 1 0 9 2020 9 17...

Page 2: ...r information replaced icons 1 2 1 Added a description of the NCC s radio wave regulations 1 2 2 Organized the layout added notices on CO2 automatic calibration 1 2 3 Added a description 4 5 Start Con...

Page 3: ...ions and sensitive places such as nursing homes and day care centers and detects various types of air pollution that may have an adverse effect on the human body UNOnext provides NFC tags which can be...

Page 4: ...will be used as the switch of the air quality indicator DIP switch No 8 must also be pushed up 5 AutoMode will turn ventilation equipment on and off depending whether the air quality exceeds the thres...

Page 5: ...a control command the user presses a physical key or uses the APP for control UNOnext will perform the last command that was sent 13 If it is used in a place of business and the owner does not want to...

Page 6: ...ply Power consumption 5W Operating temperature 0 C 50 C Operating humidity 0 RH 95 RH Non condensing Storage 20 C 70 C Protection level IP20 Certification NCC FCC IC CE JRF Accessories Wall mount x 1...

Page 7: ...9 24V DC output Switch Number Switch Definition ON OFF 1 Modbus terminal resistor 2 Modbus slave position 3 3 Modbus slave position 2 4 Modbus slave position 1 5 Modbus slave position 0 6 Maximum wind...

Page 8: ...sition of each device is 0xD0 and the slave position of each device must not be repeated In order for the central control unit to receive the correct data the second to fourth pins of the DIP switch m...

Page 9: ...tches 8 On Off Mode 3 Control single speed ventilation equipment through relay This controller supports relay output The controller can control a 12V or 24V DC relay through the relay which can be tri...

Page 10: ...for power boxes in multiple countries but the power box must be installed horizontally as shown in Figure 1 2 This controller uses a 9 24V DC power supply AC power must be converted to the required p...

Page 11: ...s product is made in Taiwan The place of origin is No 256 Yangguang St Neihu Dist Taipei City Taiwan R O C iOS APP Android APP UNOweb XI APP and website links UNOnext is equipped with wireless technol...

Page 12: ...e equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help This...

Page 13: ...13 XIV Product dimensions I Device dimensions...

Page 14: ...14 II Wall mount dimensions...

Reviews: