38
⚫
SDI-12 commands
aM5!
Water content
%vol
Using
MinWool
soil-type calibration
EC
p
(pore water
EC)
mS.m
-1
Compensated to 25°C linearly at 2%/°C,
soil parameter = 4.1
T
°C
Ɛ
r
(permittivity)
(none)
EC
b
(bulk EC)
mS.m
-1
aM6!
Water content
%vol
Using
Perlite
soil-type calibration
EC
p
(pore water
EC)
mS.m
-1
Compensated to 25°C linearly at 2%/°C,
soil parameter = 4.1
T
°C
Ɛ
r
(permittivity)
(none)
EC
b
(bulk EC)
mS.m
-1
aM7!
Blank
aM8!
Blank
aM9!
Ɛ
r
(permittivity)
(none)
EC
b
(bulk EC)
mS.m
-1
T
°C
Multi-parameter measurement sets M and M1 to M8 in the table above can
be customised to better suit your application, particularly for use with third-
party (non-Delta-T Devices) loggers and controllers.
For information on how to make these measurement customisations see the
WET150 SDI-
12 Programmer’s Guide
.