Give Each WET150 a Unique Address
⚫
19
1 Open DeltaLink Transparent Mode
You need a GP2 connected to a PC running DeltaLINK v 3.9 or
later.
With
no other sensors on the network
connect the sensor to
the GP2 SDI-12 terminal.
In DeltaLINK select
Tools
,
SDI-12
,
Transparent mode
to open
theSDI-12 Transparent Mode Terminal as shown below.
Once the SDI-12 Transparent Mode Terminal is open, the GP2
behaves like a transparent serial link between SDI-12 sensors
and the PC.
2 Find out WET150’s current address
Send a command to tell the sensor to reply with its address.
You type
?!
Reply
a
We usually expect the address of a new sensor is to be zero, but
it might not be. If you already have another sensor with the
same address on the same network the command will not work.
?!
1
2
3
4
Any logger on COM1
Example