![Delta OHM HD2817T Series Manual Download Page 53](http://html1.mh-extra.com/html/delta-ohm/hd2817t-series/hd2817t-series_manual_2478334053.webp)
53
(**) Table of status codes of RL1 and RL2 relays
Bit
7
6
5
4
3
2
1
0
Meaning
1
1
1
1
1
1
1
1
0xFF
Force energised
0
0
0
0
0
0
0
1
0x01
Active_below (low threshold)
0
0
0
0
0
0
1
0
0x02
Active_ above (high threshold)
0
0
0
0
0
0
1
1
0x03
Active_on_both (both are active)
0
0
0
0
0
0
0
0
0x00
Idle
0
0
0
0
0
1
x
x
0x04
Hysteresis
0
0
0
0
1
x
x
x
0x08
Force energised on error (please
see the item ON ERROR in the
menu)
(***) Table of status codes of RL3 relay
Bit
7
6
5
4
3
2
1
0
Meaning
0
1
x
x
x
x
x
x
Force de-energised
1
1
x
x
x
x
x
x
Force energised
0
0
x
x
x
x
x
1
Normally de-energised, energised on
probe 1 error
1
0
x
x
x
x
x
1
Normally energised, de-energised on
probe 1 error
0
0
x
x
x
x
1
0
Normally de-energised, energised on
probe 2 error
1
0
x
x
x
x
1
0
Normally energised, de-energised on
probe 2 error
0
0
x
x
x
x
1
1
Normally de-energised, energised on
probe 1 or 2 error
1
0
x
x
x
x
1
1
Normally energised, de-energised on
probe 1 or 2 error
x
x
x
x
x
1
x
x
Analog output 1 out of range
x
x
x
x
1
x
x
x
Analog output 2 out of range
x
x
x
1
x
x
x
x
Analog output 3 out of range
Summary of Contents for HD2817T Series
Page 62: ...62...