Source
It lists the alarm sources for the “
Energized on error” and “Deenergized on error”
functioning modes
.
RELAY RL3
SOURCE OF ALARM
*
Probe Error
*
Out 1 Error
Out 2 Error
*
Out 3 Error
<ESC> exit/cancel
Probe error
is triggered when a probe error occurs, e.g. when the measurement
exceeds the functioning limits indicated in the technical information, when the
probe becomes faulty or gets disconnected.
Out 1 Error
is triggered when the analog output 1 goes in error because the set high
or low limits are exceeded.
Out 2 Error
is triggered when the analog output 2 goes in error because the set high
or low limits are exceeded.
Out 3 Error
is triggered when the analog output 3 goes in error because the set high
or low limits are exceeded.
You can select multiple error sources, as illustrated in the example: the occurrence of
any of them triggers the alarm and switches the relay RL3.
An asterisk before the name indicates that the item is enabled. Use the
arrows to
invert its state and press ENTER. Repeat for all items. Press ESC to return to the
previous menu.
Delay
It represents the delay before generating an alarm.
RELAY RL3
DELAY TIME
delay: 10 (sec)
<UP> <DOWN> change
<ENTER> confirm
<ESC> exit/cancel
To modify the delay time, set a new value using the
arrows and press ENTER to
confirm. Press ESC to return to the previous menu.