44
COMMAND
ACTION
RESPONSE
NOTES
AZ
Full heading
HD2256-2
Firmware 1.00.100
2006_01_31
Ser. number=00000000
Calibrated 01-FEB-06 00:01:00
User=Administrator
Communication interface=USB
Temp.comp. mode=AUTOMATIC
Temp. sensor = Comb. Chi Pt100
DA
Input date-time
&/?
"From 2005/12/12 12:34:56"
It rejects any incorrect date
FA
Clock date request
060414092400
Current date “yy/mm/dd
hh/mm/ss”, HEX format
FB
Clock date request
01-JAN-06 00:33:27
Month name according to the
language
FD
Instrument calibration request
060414092400
Calibration date “yy/mm/dd
hh/mm/ss”, HEX format
FE
Instrument calibration request
01-JAN-06 00:00:00
Month name according to the
language
K1
Print current measurements
Same as manual print
It increases the ID
K2
Calibration status pH1
Same as manual print
K3
Calibration status pH2
Same as manual print
Only HD2205.2
K4
Calibration history pH1
Same as manual print
K5
Calibration history pH2
Same as manual print
Only HD2205.2
K6
Last conductivity calibration
Same as manual print
Only HD2206.2 and HD2256.2
KE
Exit from memory mode
&
KL
Activate log
&
KM
Activate memory display
mode
&
Not accepted if no data present in
the memory
KQ
Stop log + save operating pa-
rameters
&
Use it also to only save the oper-
ating parameters
KS
Single line continuous print-
ing
&
KT
Stop single line printing
&
LDxxxx
Memory dump no. xxxx+1
Dump or ?
LN
No. of next memory location
request
Next avail. memory=0001
LR
Setting of the displayed
memory index
&/?
lr3 ---> shows memory no. 4
P0
Ping & lock keys
&
P1
Ping & unlock keys
&
RA
Read log interval
Print Interval= 000
RE
Read current endpoint status
Endpoint mode = 0
RF
Read parameter
α
T
Chi alfa = 2.00
RH
Read pH electrode calibration
expiration
pH cal exp.days = 0
RI
Read parameter ID
Sample ID= 00000001
Summary of Contents for HD2205.2